## Abstract Polyethylene wear particles are one of the most important factors affecting the results of total hip arthroplasty. Hylamer, a highly crystalized ultra‐high‐molecular‐weight polyethylene, has been used in total hip arthroplasty for wear reduction, but has exhibited high wear rate and exc
Wear particle analysis of highly crosslinked polyethylene isolated from a failed total hip arthroplasty
✍ Scribed by Yukihide Minoda; Akio Kobayashi; Akira Sakawa; Masaharu Aihara; Koichi Tada; Ryo Sugama; Kentaro Iwakiri; Hirotsugu Ohashi; Kunio Takaoka
- Publisher
- John Wiley and Sons
- Year
- 2008
- Tongue
- English
- Weight
- 163 KB
- Volume
- 86B
- Category
- Article
- ISSN
- 1552-4973
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Polyethylene wear particles are one of the most important factors affecting the results of total hip arthroplasty (THA). To reduce wear generation and to achieve better long‐term results of THA, highly crosslinked polyethylene (HXPE) has recently been introduced and come into wide use. Thus far, however, there have been no reports on in vivo analysis of HXPE wear particles. We isolated HXPE wear particles from periprosthetic tissue of a failed THA and analyzed using scanning electron microscope. The number of particles was 5.33 × 10^7^ g^−1^. Particle size (equivalent circle diameter) was 0.66 ± 0.40 μm (mean ± standard error). Aspect ratio and roundness were 1.37 ± 0.26 and 1.44 ± 0.67, respectively. All the particles were round shaped, and “fibrils” or “shreds” were not detected. Thus far, this was the first report on in vivo wear particle analysis of HXPE. HXPE generated less, smaller, and rounder particles, compared with the corresponding reported values for particles generated from conventional polyethylene. These characteristics might affect macrophage response, osteolysis, and long‐term results of THA with HXPE. © 2008 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 2008
📜 SIMILAR VOLUMES
Periprosthetic tissue obtained at revision surgery from eight total hip replacement cases was hydrolyzed, and polyethylene debris particles were isolated from each case. Individual particles were analyzed by scanning electron microscopy (SEM) and computerized image analysis in accordance with ASTM F
## Abstract Reduction of wear with highly crosslinked polyethylene (HXLPE) has been reported in __in vitro__ and __in vivo__ studies of total hip prostheses. However, use of HXLPE in total knee prostheses is still controversial. The aim of this study was to compare __in vivo__ polyethylene wear par
## Abstract The size, shape, and number of polyethylene wear particles found in synovial fluids of patients 1 year after implantation of 22 well‐functioning total knee prostheses (11 contemporary mobile‐bearing type, 11 posterior‐stabilized type) were determined. Polyethylene wear particles were is
## Abstract Three cemented femoral components from total hip arthroplasties retrieved at autopsy were examined in regard to the migration of birefringent polyethylene particles across the fixation interfaces. They were obtained from two patients who died from unrelated causes. They had been in plac
The less intense tissue reaction around metal on metal total hip replacements (THRs) compared to metal on polyethylene (PE) THRs may be explained by the differences in the characteristics of metal wear particles. In this study, transmission electron microscopy was used to study metal wear particles