## Abstract Polyethylene wear particles are one of the most important factors affecting the results of total hip arthroplasty (THA). To reduce wear generation and to achieve better long‐term results of THA, highly crosslinked polyethylene (HXPE) has recently been introduced and come into wide use.
Standardized analysis of UHMWPE wear particles from failed total joint arthroplasties
✍ Scribed by Mabrey, Jay D. ;Afsar-Keshmiri, Armin ;Engh, Gerard A. ;Sychterz, Christi J. ;Wirth, Michael A. ;Rockwood, Charles A. ;Agrawal, C. Mauli
- Publisher
- John Wiley and Sons
- Year
- 2002
- Tongue
- English
- Weight
- 743 KB
- Volume
- 63
- Category
- Article
- ISSN
- 0021-9304
No coin nor oath required. For personal study only.
✦ Synopsis
Periprosthetic tissue obtained at revision surgery from eight total hip replacement cases was hydrolyzed, and polyethylene debris particles were isolated from each case. Individual particles were analyzed by scanning electron microscopy (SEM) and computerized image analysis in accordance with ASTM F1877-98, a standard for quantitative description of wear debris. For comparison, periprosthetic tissues from eight total knee revision and four total shoulder revision cases were processed and analyzed with identical methods. A total of 2599 hip, 4345 knee, and 1200 shoulder particles were analyzed. The morphologies of the isolated polyethylene particles from the total hip specimens were distinctly different from the total knee and total shoulder particles. The mean equivalent circle diameter (ECD) for hip particles was 0.694 microm +/- 0.005; knee particles measured 1.190 microm +/-0.009; and shoulder particles 1.183 microm +/- 0.017. The ECD was significantly different between hip particles and those from the shoulder and knee. The mean aspect ratio (AR) for the hip particles was 1.626 +/- 0.015, compared to the knee particles at 1.935 +/- 0.015 and shoulder particles at 2.082 +/- 0.033. The AR was statistically different among all three groups. Other descriptors from the ASTM standard, elongation (E), form factor (FF), and roundness (R) were all significantly different among the three groups of joints. This study demonstrates the utility of ASTM F1877-98 in differentiating wear debris particles from different sources.
📜 SIMILAR VOLUMES
## Abstract Polyethylene wear particles are one of the most important factors affecting the results of total hip arthroplasty. Hylamer, a highly crystalized ultra‐high‐molecular‐weight polyethylene, has been used in total hip arthroplasty for wear reduction, but has exhibited high wear rate and exc
## Abstract The size, shape, and number of polyethylene wear particles found in synovial fluids of patients 1 year after implantation of 22 well‐functioning total knee prostheses (11 contemporary mobile‐bearing type, 11 posterior‐stabilized type) were determined. Polyethylene wear particles were is
## Abstract It is recognized that the chronic inflammation in peri‐prosthetic tissue that contributes to implant failure frequently is provoked by the presence of wear debris. Some wear debris is inevitable because of the nature of the prosthesis, but not all patients develop severe inflammatory re
## Abstract There is currently considerable interest in the wear debris and osteolytic potential of different types of bearings used in total joint replacements. The biological activity of the wear debris is dependent on the size and volume of the particles produced. Wear volume also plays an impor