Total hip joint prosthesis for in vivo measurement of forces and moments
β Scribed by Philipp Damm; Friedmar Graichen; Antonius Rohlmann; Alwina Bender; Georg Bergmann
- Publisher
- Elsevier Science
- Year
- 2010
- Tongue
- English
- Weight
- 679 KB
- Volume
- 32
- Category
- Article
- ISSN
- 1350-4533
No coin nor oath required. For personal study only.
β¦ Synopsis
A new instrumented hip joint prosthesis was developed which allows the in vivo measurement of the complete contact loads in the joint, i.e. 3 force and 3 moment components. A clinically proven standard implant was modified. Inside the hollow neck, 6 semiconductor strain gauges are applied to measure the deformation of the neck. Also integrated are a small coil for the inductive power supply and a 9-channel telemetry transmitter. The neck cavity is closed by a titanium plate and hermetically sealed by electron beam welding. The sensor signals are pulse interval modulated (PIM) with a sampling rate of about 120 Hz. The pulses are transmitted at radio frequencies via a small antenna loop inside the ceramic head, which is connected to the electronic circuit by a two-pin feedthrough. Inductive power supply, calculation of the loads from the measured deformations and real time load display are carried out by the external equipment. The maximum error of the load components is 2% including crosstalk.
π SIMILAR VOLUMES
A three-dimensional computer model of the pelvis, femur, gluteus medius, and gluteus minimus was used to evaluate the changes in muscle moment arms and force-generating capacities caused by alterations in the location of the greater trochanter. In the first part of this study, the hip center and all
## Abstract We introduce here a technique to measure the threeβdimensional kinematics and laxity characteristics of the ankle joint complex __in vivo__. The system consists of an optoelectric, kinematic data acquisition system that is used to measure the motion of the ankle joint complex in respons
## Abstract The purpose of this study was to determine the effects of superior displacement of the hip center and changes in three prosthetic parameters (neck length, neckβstem angle, and anteversion angle) on the capacity of muscles to generate force and moment about the hip. A threeβdimensional m