Background and Objectives: The high-power Thulium fiber laser has previously been shown to rapidly vaporize and coagulate soft urinary tissues (e.g., prostate). This is the first preliminary study of a high-power Thulium fiber laser for fragmentation of urinary stones. Study Design/Materials and Met
Thulium fiber laser lithotripsy using tapered fibers
✍ Scribed by Richard L. Blackmon; Pierce B. Irby; Nathaniel M. Fried
- Book ID
- 102471232
- Publisher
- John Wiley and Sons
- Year
- 2010
- Tongue
- English
- Weight
- 140 KB
- Volume
- 42
- Category
- Article
- ISSN
- 0196-8092
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Introduction
The Thulium fiber laser has recently been tested as a potential alternative to the Holmium:YAG laser for lithotripsy. This study explores use of a short taper for expanding the Thulium fiber laser beam at the distal tip of a small‐core fiber.
Methods
Thulium fiber laser radiation with a wavelength of 1,908 nm, 10 Hz pulse rate, 70 mJ pulse energy, and 1‐millisecond pulse duration was delivered through a 2‐m‐length fiber with 150‐µm‐core‐input‐end, 300‐µm‐core‐output‐end, and 5‐mm‐length taper, in contact with human uric acid (UA) and calcium oxalate monohydrate (COM) stones, ex vivo (n = 10 each). Stone mass loss, stone crater depths, fiber transmission losses, fiber burn‐back, irrigation rates, and deflection through a flexible ureteroscope were measured for the tapered fiber and compared with conventional fibers.
Results
After delivery of 1,800 pulses through the tapered fiber, mass loss measured 12.7±2.6 mg for UA and 7.2±0.8 mg COM stones, comparable to conventional 100‐µm‐core fibers (12.6±2.5 mg for UA and 6.8±1.7 mg for COM stones). No transmission losses or burn‐back occurred for the tapered fiber after 36,000 pulses, while a conventional 150‐µm fiber experienced significant tip degradation after only 1,800 pulses. High irrigation rates were measured with the tapered fiber inserted through the working port of a flexible ureteroscope without hindering its deflection, mimicking that of a conventional 150 µm fiber.
Conclusions
The short tapered distal fiber tip allows expansion of the laser beam, resulting in decreased fiber tip damage compared to conventional small‐core fibers, without compromising fiber bending, stone vaporization efficiency, or irrigation rates. Lasers Surg. Med. 42:45–50, 2010. © 2010 Wiley‐Liss, Inc.