Abstract

Introduction

The holmium:YAG laser is currently the most common laser lithotripter. However, recent experimental studies have demonstrated that the thulium fiber laser is also capable of vaporizing urinary stones. The high‐temperature water absorption coefficient for the thulium wavelength (µ~a~ = 160 cm^−1^ at λ = 1,908 nm) is significantly higher than for the holmium wavelength (µ~a~ = 28 cm^−1^ at λ = 2,120 nm). We hypothesize that this should translate into more efficient laser lithotripsy using the thulium fiber laser. This study directly compares stone vaporization rates for holmium and thulium fiber lasers.

Methods

Holmium laser radiation pulsed at 3 Hz with 70 mJ pulse energy and 220 microseconds pulse duration was delivered through a 100‐µm‐core silica fiber to human uric acid (UA) and calcium oxalate monohydrate (COM) stones, ex vivo (n = 10 each). Thulium fiber laser radiation pulsed at 10 Hz with 70 mJ pulse energy and 1‐millisecond pulse duration was also delivered through a 100‐µm fiber for the same sets of 10 stones each.

Results

For the same number of pulses and total energy (126 J) delivered to each stone, the mass loss averaged 2.4±0.6 mg (UA) and 0.7±0.2 mg (COM) for the holmium laser and 12.6±2.5 mg (UA) and 6.8±1.7 (COM) for the thulium fiber laser.

Conclusions

UA and COM stone vaporization rates for the thulium fiber laser averaged 5–10 times higher than for the holmium laser at 70 mJ pulse energies. With further development, the thulium fiber laser may represent an alternative to the conventional holmium laser for more efficient laser lithotripsy. Lasers Surg. Med. 42:232–236, 2010. © 2010 Wiley‐Liss, Inc.