๐”– Bobbio Scriptorium
โœฆ   LIBER   โœฆ

The use of silicon in the preparation of apolar glass capillary columns

โœ Scribed by Bertsch, W. ;Pretorius, V. ;van Niekerk, G. C.


Publisher
John Wiley and Sons
Year
1982
Tongue
English
Weight
526 KB
Volume
5
Category
Article
ISSN
0935-6304

No coin nor oath required. For personal study only.

โœฆ Synopsis


Abstract

As an extention of previous reports, variables affecting the performance of columns modified by silicon deposition have been optimized. The aim of this work was to produce thermostable apolar glass capillary columns regardless of the glass type used and to maintain column performance under stressful conditions. The main parameters studied were the influence of glass type, leaching effects, and silicon layer thickness on column activity. Deactivation and stationary phase conditions were held constant. Both coated and uncoated columns were tested.

The thickness of the silicon layer was found to be relatively unimportant. There was no difference whether soft glass or borosiiicate glass was used and leaching before silicon deposition did not influence column activity. Bare silicon surfaces showed considerable activity especially in respect to interactions with free acids and bases.

There are indications that the surface consists of silicon oxide and other oxygenated forms of silicon rather than of the element. Treatment of the silicon layers with dilute hydrofluoric acid and the strict exclusion of traces of oxygen and water did not improve the situation. In spite of such specific interactions, silicon surfaces were easily deactivated by heat treatment with polysiloxanes.

Silicon surfaces deactivated by baking with octamethyltetrasiloxane at 400ยฐC are inert and temperature stable and show characteristics similar to persilylated surfaces.


๐Ÿ“œ SIMILAR VOLUMES


Preparation of apolar glass capillary co
โœ Grob, K. ;Grob, G. ;Grob, K. ๐Ÿ“‚ Article ๐Ÿ“… 1978 ๐Ÿ› John Wiley and Sons ๐ŸŒ English โš– 520 KB

## Abstract As an extension of previous reports, the barium carbonate procedure has been optimized in detail for the preparation of apolar columns. The aim was to produce optimum overall column characteristics, and to maintain them unchanged under the prolonged influence of the highest possible tem

A warning against the use of diluted sto
โœ Grob, K. ;Grob, G. ๐Ÿ“‚ Article ๐Ÿ“… 1978 ๐Ÿ› John Wiley and Sons ๐ŸŒ English โš– 174 KB ๐Ÿ‘ 1 views

For the preparation of capillary columns by the static coating method [I] considerable volumes of solution containing the liquid phase in a concentration range of 0.05-2.0% are used. For convenience and economy we have for many years prepared such solutions in batches of 50-100 ml, and have continue

The preparation of thick-film glass capi
โœ Torline, P. ;du Plessis, G. ;Schnautz, N. ;Thompson, J. C. ๐Ÿ“‚ Article ๐Ÿ“… 1979 ๐Ÿ› John Wiley and Sons ๐ŸŒ English โš– 354 KB ๐Ÿ‘ 1 views

## Abstract A detailed method for the routine preparation of glass capillary columns is presented. The method consists of coating a glass tube with quartz powder prior to pulling the tube into a capillary. The inner surface of the capillary consists of an even distribution of quartz particles fused