The influence of material and design features on the mechanical properties of transpedicular spinal fixation implants

✍ Scribed by Hahn, Michael ;Nassutt, Roman ;Delling, G�nter ;Mahrenholtz, Oskar ;Schneider, Erich ;Morlock, Michael


Publisher
John Wiley and Sons
Year
2002
Tongue
English
Weight
965 KB
Volume
63
Category
Article
ISSN
0021-9304

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

This study describes the finding and performance of mechanical strength and corrosion testing procedures for comparative examination of multiple internal transpedicular spine fixators. Seven different implant models from five different manufacturers were compared regarding their bending strength and fatigue resistance. Because of the unacceptably high levels of time and material that they require, ISO and ASTM testing standards are not applicable to comparative testing. In addition, there is a lack of knowledge about clinically defined and proven strength‐limit values. Therefore, actual standard testing procedures have been modified and extended to corrosion testing. Overall, the effort necessary to obtain reproducible comparative data has been reduced significantly. Although a reduced number of implants of each type were available for destructive testing, the results revealed fundamental differences in the tested implants between different materials and design features. During fatigue testing some of the implants showed poor corrosion properties. Because spinal fixation implants tend to be used as long‐duration implants, corrosion testing as well as comparative strength testing with clinically successful implant models should be performed as preclinical evaluations. © 2002 Wiley Periodicals, Inc. J Biomed Mater Res (Appl Biomater) 63: 354–362, 2002


📜 SIMILAR VOLUMES


The influence of fixation peg design on
✍ N. J. Giori; G. S. Beaupré; D. R. Carter 📂 Article 📅 1990 🏛 Elsevier Science 🌐 English ⚖ 694 KB

## Abstract The variety of fixation peg designs existing on prosthetic implants indicates uncertainty regarding the optimum design of fixation pegs for the reduction of stress and relative motion at the bone‐implant interface. Fixation pegs have a number of important functions on a prosthesis, one

Influence of stability and mechanical pr
✍ Mochida, Yuichi ;Bauer, Thomas W. ;Nitto, Hironori ;Kambic, Helen E. ;Muschler, 📂 Article 📅 2000 🏛 John Wiley and Sons 🌐 English ⚖ 176 KB

A prospective and quantitative animal study was performed to evaluate the production of wear particles from a spinal fixation device, and to test the hypothesis that the concentration of wear debris particles adjacent to spinal fixation hardware is correlated with the stiffness of the spinal fusion