## Abstract Hip resurfacing arthroplasty has become a popular alternative to conventional hip surgery. Surface replacements with bearing sizes of 55 mm (__n__ = 5) and 39 mm (__n__ = 5) were tested in a hip simulator for 15 million cycles (Mc). Wear debris was isolated from the serum lubricant and
The influence of clearance on friction, lubrication and squeaking in large diameter metal-on-metal hip replacements
✍ Scribed by Claire L. Brockett; Phil Harper; Sophie Williams; Graham H. Isaac; Rob S. Dwyer-Joyce; Zhongmin Jin; John Fisher
- Publisher
- Springer
- Year
- 2007
- Tongue
- English
- Weight
- 231 KB
- Volume
- 19
- Category
- Article
- ISSN
- 0957-4530
No coin nor oath required. For personal study only.
📜 SIMILAR VOLUMES
## Abstract This article aims to clarify the influence of design‐ and manufacturing‐related parameters on wear of metal‐on‐metal (MoM) joint bearings. A database search for publications on wear simulator studies of MoM bearings was performed. The results of published studies were normalized; groups