✦ Synopsis


We have assessed the performance of a series of ventilators (modified versions of the ventiPAC, paraPAC and babyPAC ventilators; SIMS pneuPAC Ltd, Luton, UK) in a magnetic resonance imaging (MRI) scanning environment, with MR safety and compatibility issues being addressed. Following initial modifications to remove ferromagnetic components and replace them with MR‐safe materials, all three ventilators performed well in a series of tests in static magnetic fields up to 2 T. Ventilator performance was unaffected by static fields, switching gradients or radio frequency fields within the MR suite. Furthermore, the devices produced no degradation of image quality when used during MR scanning. We discuss management strategies for the care of critically ill ventilated patients during MR procedures.


📜 SIMILAR VOLUMES


A new perfluorocarbon for use in fluorin
✍ Christopher H. Sotak; Paul S. Hees; Hsu-Nan Huang; Ming-H. Hung; Carl G. Krespan 📂 Article 📅 1993 🏛 John Wiley and Sons 🌐 English ⚖ 874 KB

## Abstract A new perfluorocarbon, PTBD (perfluoro‐2,2, 2′, 2′‐tetrameth‐yl‐4,4′‐bis(1, 3‐dioxolane)), is described for use in ^19^F MR imaging and spectroscopy. Two‐thirds of the molecular fluorine in PTBD resonates at a single frequency and can be imaged without the use of frequency‐selective spi

Use of high temperature superconductor i
✍ A. S. Hall; N. Mcn. Alford; T. W. Button; D. J. Gilderdale; K. A. Gehring; I. R. 📂 Article 📅 1991 🏛 John Wiley and Sons 🌐 English ⚖ 266 KB 👁 1 views

## Abstract A planar YBa~2~Cu~3~O~x~ (YBCO) coil has been used as a pick‐up coil in a nuclear magnetic resonance imager. It was found that the quality factor __Q__ of the YBCO coil was greater than its copper counterpart by approaching 50% even in a field of 0.15 T. The Q of the copper coil at 77 K

Development, standardization, and testin
✍ Debra M. Ikeda; Nola M. Hylton; Karen Kinkel; Mary G. Hochman; Christiane K. Kuh 📂 Article 📅 2001 🏛 John Wiley and Sons 🌐 English ⚖ 73 KB 👁 1 views

## Abstract The purpose of this study was to develop, standardize, and test reproducibility of a lexicon for reporting contrast‐enhanced breast magnetic resonance imaging (MRI) examinations. To standardize breast MRI lesion description and reporting, seven radiologists with extensive breast MRI exp