๐”– Bobbio Scriptorium
โœฆ   LIBER   โœฆ

Temperature and pressure effects during erbium laser stapedotomy

โœ Scribed by Pratisto, Hans; Frenz, Martin; Ith, Michael; Romano, Valerio; Felix, Dominik; Grossenbacher, Rudolf; Altermatt, Hans J.; Weber, Heinz P.


Publisher
John Wiley and Sons
Year
1996
Tongue
English
Weight
934 KB
Volume
18
Category
Article
ISSN
0196-8092

No coin nor oath required. For personal study only.

โœฆ Synopsis


Background and Objectives: Laser-assisted stapedotomy has become a well-established alternative to the mechanical drilling method. The goal of this study is to quantify the mechanical and thermal tissue effects and to determine optimum erbium laser parameters for safe clinical treatment. Study DesignlMaterials and Methods: On an inner ear model, time-resolved pressure measurements and Schlieren optical flash photography were performed during the perforation of the stapes foot plate using an erbium laser at 2.79 pm. The laser radiation was transmitted via an optical zirconium fluoride fiber. The laser-treated foot plates were investigated by light microscopy and scanning electron microscopy to visualise the laserinduced tissue effects. Results: Perforation of the stapes foot plate can be performed with a few erbium laser pulses with high precision and a thermal damage zone of <lo pm. Strong pressure transients were found to be generated by the bone ablation process and the collapse of a vapor channel created in the perilymph after fenestration.

Conclusion:

From the comparison of the laser-induced pressure with the limit graph to avoid hearing defects published by Pfander, an unobjectionable use of the erbium laser is deduced for fluences <10 J/cm2. The erbium laser seems to represent an ideal instrument for middle ear surgery with all the advantages (precision, fiber optic transportable, high ablation efficiency, safety) desired for clinical application. o 19% Wiley-Liss, Inc.


๐Ÿ“œ SIMILAR VOLUMES


Thermic effects in the โ€œvestibuleโ€ durin
โœ Jovanovic, S.; Schรถnfeld, U.; Fischer, R.; Dรถring, M.; Prapavat, V.; Mรผller, G.; ๐Ÿ“‚ Article ๐Ÿ“… 1998 ๐Ÿ› John Wiley and Sons ๐ŸŒ English โš– 253 KB ๐Ÿ‘ 1 views

## Background and Objective : Apart from the ablation properties at the stapes footplate, the degree of thermic loading in the inner ear is important in determining the suitability of pulsed lasers for stapedotomy. The aim of the study is to compare the thermic effects in the vestibule with differ

Switchable dual wavelength erbium-doped
โœ Young-Geun Han; Ju Han Lee ๐Ÿ“‚ Article ๐Ÿ“… 2007 ๐Ÿ› John Wiley and Sons ๐ŸŒ English โš– 123 KB

## Abstract A simple technique of a switchable dual wavelength erbiumโ€doped fiber laser based on fiber Bragg gratings is proposed and demonstrated by using both the linear cavity loss controlling and the homogeneous gain broadening of EDF. The single or dual lasing wavelength can be also obtained b

Room-temperature multiwavelength operati
โœ Gautam Das; J. W. Y. Lit ๐Ÿ“‚ Article ๐Ÿ“… 2002 ๐Ÿ› John Wiley and Sons ๐ŸŒ English โš– 128 KB

## Abstract A simple and efficient technique to achieve stable multiwavelength laser operation at room temperatures with the use of an ellipticalโ€core erbiumโ€doped fiber is proposed and demonstrated. The results give a maximum of five simultaneous laser lines in the L band. A polarization controlle