๐”– Bobbio Scriptorium
โœฆ   LIBER   โœฆ

Selective determination of oxygenates in complex samples with the O-FID analyzer

โœ Scribed by Verga, G. R. ;Sironi, A. ;Schneider, W. ;Frohne, J. Ch.


Book ID
102350385
Publisher
John Wiley and Sons
Year
1988
Tongue
English
Weight
412 KB
Volume
11
Category
Article
ISSN
0935-6304

No coin nor oath required. For personal study only.

โœฆ Synopsis


A gas-chromatographic analyzer allowing the selective detection of individual oxygenates in complex organic mixtures, such as gasolines, is described. The analyzer is based on the oxygenspecific response flame ionization detection (0-FID) method. The system operates with capillary columns and includes a cracking reactor to convert any oxygenate to carbon monoxide and a special FID equipped with a microreactor for the catalytic hydrogenation of CO and detection as methane. Hydrocarbons give no signal. The selectivity of the method is better than 1:107 and the linear range approaches lo5. The analysis of modern fuels containing oxygenated additives, to improve octane ratings, may greatly benefit from the proposed analytical method.