Relative ion yields measured with a high-resolution glow discharge mass spectrometer operated with an argon/hydrogen mixture

โœ Scribed by R.W. Smithwick III; D.W. Lynch; J.C. Franklin


Book ID
103995770
Publisher
Elsevier Science
Year
1993
Tongue
English
Weight
722 KB
Volume
4
Category
Article
ISSN
1044-0305

No coin nor oath required. For personal study only.

โœฆ Synopsis


Quantitative elemental analysis by glow discharge mass spectrometry (GDMS) requires a calibration factor for each element. The calibration factors used in the present work are called relative ion yields (RIYs). The RIYs of each of 19 elements within samples of four National Institute of Standards and Technology steel reference materials (nos. 661-664) were measured using pure argon and an argon mixture containing 1.0% hydrogen by volume. The RIYs measured using pure argon correlated within a factor of approximately 2-3 to the RIYs calculated by a theoretical model. The RlYs measured for these 19 elements using the argon mixture containing 1.0% hydrogen correlated within a factor of approximately 1.3 to the calculated RIYs. These results may have significant analytical potential with respect to GDMS and may have application to other plasma techniques.