𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Reduction of fretting corrosion of Ti-6Al-4V by various surface treatments

✍ Scribed by Anne M. Maurer; Dr. Stanley A. Brown; Joe H. Payer; Katharine Merritt; Jill S. Kawalec


Publisher
Elsevier Science
Year
1993
Tongue
English
Weight
897 KB
Volume
11
Category
Article
ISSN
0736-0266

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

Titanium and titanium‐6% aluminum‐4% vanadium (Ti‐6Al‐4V) are known to be biocompatible and corrosion resistant. However, there have been numerous reports of elevated tissue levels of titanium due to passive dissolution, wear, or fretting corrosion of implants. Studies were undertaken to determine whether the fretting corrosion of Ti‐6Al‐4V could be reduced by surface treatment of one or both surfaces in a fretting situation. Three different surface treatments were studied: ion implantation, physical vapor deposition nitriding, and plasma ion nitriding. The specimens used were screws fretting against the countersinks of a two‐hole plate. Fretting corrosion was assessed by weight loss, by chemical analysis of test solutions, and by scanning electron microscopy. Surface treatment of one component, the screws, resulted in reduction in the release of titanium to only 18–32% of that seen with the untreated controls. Weight loss of the untreated plates fretted against physical vapor deposition nitrided screws and plasma ion nitrided screws was reduced to 31 and 38% of the control, respectively. The weight loss of plasma nitrided screws was only 30% that of the control. Nitriding of both plates and screws resulted in a further decrease in plate weight loss and metal release. Plasma ion nitriding of both components had the most significant effect, with the weight loss and titanium release being only 11 and 2% of the control values, respectively.


πŸ“œ SIMILAR VOLUMES


Effect of surface treatments on fretting
✍ Patrick J. Golden; Alisha Hutson; Vasan Sundaram; James H. Arps πŸ“‚ Article πŸ“… 2007 πŸ› Elsevier Science 🌐 English βš– 351 KB

The objective of this work was to investigate the performance of several coatings and surface treatments subjected to fretting fatigue with Ti-6Al-4V specimens and pads. Five coating systems were initially chosen and tested using two different fretting fatigue apparatus. The five coating systems con

The effect of surface treatments on the
✍ Matteo Dalmiglio; Petra Schaaff; Uwe Holzwarth; Roberto Chiesa; Gianni Rondelli πŸ“‚ Article πŸ“… 2008 πŸ› John Wiley and Sons 🌐 English βš– 174 KB πŸ‘ 2 views

## Abstract Stem modularity in total hip replacement introduces an additional taper joint between Ti‐6Al‐4V stem components with the potential for fretting corrosion processes. One possible way to reduce the susceptibility of the Ti‐6Al‐4V/Ti‐6Al‐4V interface to fretting is the surface modification

Effects of precoating surface treatments
✍ Dr. Alan W. Eberhardt; Byeong S. Kim; E. Douglas Rigney; Gerald L. Kutner; Charl πŸ“‚ Article πŸ“… 1995 πŸ› Wiley (John Wiley & Sons) 🌐 English βš– 454 KB πŸ‘ 2 views

## Abstract Grit blasting is a common procedure of roughening surfaces to promote physical attachment of porous coatings, but it has been shown to reduce fatigue strength. Shot peening is known to increase fatigue strength by inducing compressive surface stresses; however, it is not known how subse

Effect of ion implantation on fatigue, f
✍ S Saritas; R.P.M Procter; W.A Grant πŸ“‚ Article πŸ“… 1989 πŸ› Elsevier Science 🌐 English βš– 883 KB

Earlier studies of the effect of ion implantation on wear of Ti-6Al-4V have now been extended to an examination of the effect of ion implantation on fatigue, fretting and fretting-corrosion of this alloy. Using smooth rotating-bending fatigue specimens tested in air at a frequency of 47Hz it has bee

Mixed-metal fretting corrosion of Ti6Al4
✍ Kawalec, Jill S. ;Brown, Stanley A. ;Payer, Joe H. ;Merritt, Katharine πŸ“‚ Article πŸ“… 1995 πŸ› John Wiley and Sons 🌐 English βš– 802 KB

## Abstract Corrosion has been reported at the modular interfaces of total joint replacement implants, but with conflicting theories as to the cause of such damage. The modular design itself leaves the interface susceptible to galvanic, crevice, or fretting corrosion, or a combination of the three.