𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Quantitative gas-liquid chromatographic determination of ftorafur and 5-fluorouracil in biological specimens

✍ Scribed by C. Pantarotto; R. Fanelli; S. Filippeschi; T. Facchinetti; F. Spreafico; M. Salmona


Publisher
Elsevier Science
Year
1979
Tongue
English
Weight
413 KB
Volume
97
Category
Article
ISSN
0003-2697

No coin nor oath required. For personal study only.

✦ Synopsis


A method developed for measuring the antitumor agent ftorafur and its biotransformation product Muorouracil was applied to biological specimens. After extraction with ethylacetate, ftorafur, S-fluorouracil, and the internal standard 2-methyl-4-hydroxy-6-chloromethylpyrimidine are converted to their chloromethyldimethylsilyl derivatives and assayed by glc, using either an electron-capture or a flame ionization detector. The minimum detectable amount is 100 pg/injection for ftorafur and 50 pg/injection for 5fluorouracil employing electron-capture detection. Linearity was found up to microgram amounts of both substances, without any interference from endogenous substrates. Preliminary data are reported on the comparative serum kinetics of ftorafur and 5-fluorouracil in mice.