β¦ Synopsis
Traces of aluminium were preconcentrated in water samples by forming a complex with lumogallion, and successfully accumulated in the micellar phase obtained from cationic or non-ionic surfactants and filtered through 10000 molecular weight cut-off ultrafiltration membranes. Studies indicated that, at pH 5.9, with 1 x 10m3 M lumogallion and 2 X lo-* M cetyltrimethylamonium bromide, quantitative retention of aluminium present at pg ml-' concentration levels was achieved. A micellar-enhanced fluorimetric procedure was employed to analyse the permeate solutions, whereas nitrous oxide-acetylene flame atomic emission spectrometry was used to determine a1uminium in the retentate. The experimental parameters that influence the micellar-enhanced ultrafiltration (MEUF) efficiency were investigated and the analytical procedures were optimized. The MEUF of trace levels of aluminium allows for either preconcentration of the analyte or enhancement of the emission signals.
π SIMILAR VOLUMES
The ultrafiltration of micellar solutions containing nitrobenzene and 4-nitrophenol using three types of polymeric membranes was studied. Hexadecyltrimethylammonium bromide (CTAB), oxyethylated coconut fatty acid methyl esters of an average oxyethylation degree equal to 10 (OMC-10) and oxyethylated