✦ Synopsis


Space position and orientation of the axis of solenoidal magnet in space coordinates connected with warm hole cryostat body may be matter of interest. Fast and precise method was developed based on mapping the field vector component B on parallel circular paths of common axis under and over the magnet field center. Coincidence of this geometrical axis connected with upper and lower flanges of cryostat with magnetic one results in rotational independence of B values i.e. B(~o) = coast. Any mis aligninent results in quasi harmonic behaviour B(90) with amplitude which depends on degree of mis alignment. lterative procedure reducing value of this mis alignment amplitude up to regaining prescribed degree of axes coincidence and all the device are described.