Photocured polymers in ion-selective electrode membranes Part 6: Photopolymerized lithium sensitive ion-selective electrodes for flow injection potentiometry

โœ Scribed by J.R. Farrell; P.J. Iles; T. Dimitrakopoulos


Publisher
Elsevier Science
Year
1996
Tongue
English
Weight
515 KB
Volume
335
Category
Article
ISSN
0003-2670

No coin nor oath required. For personal study only.

โœฆ Synopsis


A photocured lithium ion-selective electrode, based on the ionophore N,N-dicyclohexyl-N',N'-diisobutyl-cis-cyclohexane-1,2-dicarboxamide (ETH 1810), was developed and evaluated. The robust nature of the photocured membrane made it ideally suitable for measurements in flow injection potentiometry within a linear range 0.1-10-3 M, detection limit of 5 x 10e4 M, and sample throughput of 150 h-l. The electrode was used successfully to determine lithium levels in pharmaceutical lithium carbonate tablets.


๐Ÿ“œ SIMILAR VOLUMES


Photocured polymers in ion-selective ele
โœ J.R. Farrell; P.J. Iles; T. Dimitrakopoulos ๐Ÿ“‚ Article ๐Ÿ“… 1996 ๐Ÿ› Elsevier Science ๐ŸŒ English โš– 378 KB

A photocured sodium ion-selective electrode, based on the ionophore N,N-dibenzyl-N,N'-diphenyl-1,2-phenylenedioxydiacetamide (ETH 157), was developed and evaluated. The robust nature of the photocured membrane made it ideally suited for measurements in flow injection potentiometry. The detection lim

Photo-cured polymers in ion-selective el
โœ T.J. Cardwell; R.W. Cattrall; P.J. Iles; I.C. Hamilton ๐Ÿ“‚ Article ๐Ÿ“… 1988 ๐Ÿ› Elsevier Science ๐ŸŒ English โš– 245 KB

The preparation of a potassium-selective electrode based on a photo-cured polymer membrane containing valinomycin is reported. This electrode has similar response characteristics to one based on PVC with the same neutral carrier. It is suitable for use in flow-injection systems because of its fast i

Evaluation of a new wall-jet flow-throug
โœ Lucy Tina Di Benedetto; Telis Dimitrakopoulos ๐Ÿ“‚ Article ๐Ÿ“… 1997 ๐Ÿ› John Wiley and Sons ๐ŸŒ English โš– 425 KB ๐Ÿ‘ 1 views

An improved wall-jet configuration type flow-through cell has been developed for use with commercial ion-selective electrodes (ISEs) and has been evaluated in the flow injection potentiometric (FIP) mode. The flow-through cell was designed to accommodate commercial ISEs that employ a range of sensin