A simple packed-capillary high performance liquid chromatography (HPLC) system for the analysis of trace (picogram) mixtures by electrospray mass spectrometry is described in detail. This simple and cost-effective design is constructed from conventional hardware generally available in the modern LC/
Packed capillary column liquid chromatography coupled to mass spectrometry by a simple modification of an electrospray interface
✍ Scribed by Benita Hyllbrant; Niklas Tyrefors; Bengt Långström; Karin E. Markides
- Publisher
- John Wiley and Sons
- Year
- 1999
- Tongue
- English
- Weight
- 89 KB
- Volume
- 11
- Category
- Article
- ISSN
- 1040-7685
No coin nor oath required. For personal study only.
✦ Synopsis
When packed capillary columns are coupled to electrospray mass spectrometry, care must be taken to reduce the dead volumes of the detector connections. In this work, the coupling between the column and the detector was made with a 20-m i.d. fused silica capillary which was inserted in the electrospray steel needle of the probe for pneumatically assisted electrospray. Electrical contact was made at the tip of the surrounding steel needle where the liquid was eluting to the spray. The probe was then used in pure electrospray mode by turning off the assisting nitrogen gas flow. Thus, a few simple modifications of a commercial interface for pneumatically assisted electrospray made it compatible with the low flow rates from packed capillary columns. The analysis could be performed with high separation efficiency due to the low-dead-volume detector coupling. ᮊ 1999
📜 SIMILAR VOLUMES
An on-line method has been developed for the derivatization and coupled liquid chromatography (LC)/electrospray ionization (ESI) MS analysis of peptides at the femtomol level. Peptides are reacted with N-succinimidyl-2(3-pyridyl)acetate (SPA) in buffered aqueous medium at pH 7 following loading on a
A piezoelectric flow-through microdispenser interfacing capillary liquid chromatography (LC) with matrixassisted laser desorption/ionization time-of-fight mass spectrometry (MALDI-TOF MS) was developed for the identification of biomolecules. The MALDI target plate was placed on a computer controlled