A molecule ion reactor (MIR), i.e. a gas filled radio-frequency only quadrupole with a longitudinal electrical field (RFQLEF), is used as an atmospheric pressure ionization interface for an orthogonal time-of-flight mass spectrometer (O-TOFMS). A new phenomenon of selective ion 'heating' in a MIR ne
Optimization of an ion-to-photon detector for large molecules in mass spectrometry
β Scribed by F. Dubois; R. Knochenmuss; R. Zenobi
- Publisher
- John Wiley and Sons
- Year
- 1999
- Tongue
- English
- Weight
- 154 KB
- Volume
- 13
- Category
- Article
- ISSN
- 0951-4198
No coin nor oath required. For personal study only.
β¦ Synopsis
Ion packets can be detected in time-of-flight mass spectrometry by collecting the photons that are produced during the impact of the packets with a scintillator. The photon yield is a function of the ion energy. It was found that post-acceleration of the particles in front of the scintillator was an efficient way of increasing signal intensities. For the same total ion energy, the intensities were larger with post-acceleration than when only increasing the initial ion kinetic energy. A venetian blind dynode, converting the primary ion beam into electrons/secondary ions, was also introduced. Positive or negative secondary particles produced on the dynode surface could be accelerated to the scintillator. Electrons were found to give the highest signals.
Intensities similar to those measured with microchannel plates were found. The linearity and onset of saturation of the microchannel plates and the ion-to-photon detector were compared. At optimum operating conditions, the ion-to-photon detector gave around 10 times higher signals than the microchannel plates for heavy ions (150 kDa), with similar mass resolution.
π SIMILAR VOLUMES
An octupole radio frequency (rf) ion guide was evaluated for storage and filtering of ions generated by electrospray prior to introduction into an ion trap mass spectrometer. The control of the rf potential on the ion guide enabled its operation as a high-pass mass filter, removing low-mass chemical
The potential of the ion trap mass spectrometer to characterize isomeric long-chain conjugated mono-or diethylenic epoxides was investigated using the multi-stage tandem mass spectrometric (MS n ) facilities of the instrument. In the proposed approach, mass-selected reagent ions obtained from common