## Abstract A simple, specific and sensitive HPLC method with UV detection was developed and validated for the determination of tectoridin in rat plasma for the first time. Chromatographic separation was performed on a Welchrom^TM^ C~18~ column (150 × 4.6 mm, i.d., 5 µm) at a flow rate of 1.0 mL mi
On-line high performance liquid chromatography — multidimensional gas chromatography and its application to the determination of stilbene hormones in corned beef
✍ Scribed by Chappell, Colin G. ;Creaser, Colin S. ;Shepherd, Martin J.
- Publisher
- John Wiley and Sons
- Year
- 1993
- Tongue
- English
- Weight
- 425 KB
- Volume
- 16
- Category
- Article
- ISSN
- 0935-6304
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
The combination of high performance liquid chromatography interfaced on‐line with multidimensional gas chromatography (HPLC–GC–GC) is described. The HPLC column was interfaced to the GC via an on column interface, with automated pneumatic control of solvent evaporation and GC column switching. Cryogenic cold trapping was used for analyte focusing at the head of the first, non‐polar GC capillary column and optionally at the head of the second, polar column. The determination of stilbene hormones in corned beef as their methylated derivatives by flame ionization detection is described.