𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Novel tandem quadrupole-acceleration-deceleration mass spectrometer for neutralization-reionization studies

✍ Scribed by Frantisek Turecek; Ming Gu; Scott A. Shaffer


Book ID
103995574
Publisher
Elsevier Science
Year
1992
Tongue
English
Weight
897 KB
Volume
3
Category
Article
ISSN
1044-0305

No coin nor oath required. For personal study only.

✦ Synopsis


A new tandem mass spectrometer of the quadrupole-acceleration lens-deceleration. lens-quadrupole (QADQ) configuration is described. The instrument is designed for neutralization-reionization studies and consists of a 2000-u quadrupole mass analyzer as MS-I, an acceleration electrostatic lens, a series of three differentially pumped collision cells, and an electrostatic deceleration lens, energy filter, and another 2000-u quadrupole mass analyzer as MS-II. The ion optical system achieves high total ion transmission for 5-9-keV ions. Unit mass resolution in neutralization-reionization mass spectra of aromatic compounds is demonstrated. Mass, kinetic energy, and linked scans at various levels of mass resolution and sensitivity are described.


πŸ“œ SIMILAR VOLUMES


High Sensitivity Collisionally-activated
✍ Howard R. Morris; Thanai Paxton; Anne Dell; Jean Langhorne; Matthias Berg; Rober πŸ“‚ Article πŸ“… 1996 πŸ› John Wiley and Sons 🌐 English βš– 627 KB

Consideration of the special pmblems encountend in ultra-high sensitivity biopolymer sequencing studies has led to the development of a novel quadrupde/orthogonal-aderation time-of-jlight tandem mass spectrometer dcscn'bed for the first time here. The performance characteristics of this new geometry