Measurement of onset potentials for proton transfer reactions with a triple quadrupole instrument
β Scribed by Maurice M. Bursey; Thomas R. Blackburn; William J. Meyerhoffer; William A. Mattson
- Publisher
- John Wiley and Sons
- Year
- 1990
- Tongue
- English
- Weight
- 386 KB
- Volume
- 25
- Category
- Article
- ISSN
- 1076-5174
No coin nor oath required. For personal study only.
β¦ Synopsis
Problems with the treatment of data from the study of both exothermic and endothermic ion-molecule reactions are discussed. Recommendations are (i) to extrapolate data from an obviously linear region in the plot of ratios of secondary to primary ion intensity which appears -1-3 V above the value of the onset potential; (ii) to use ionization methods that give low internal energies to primary ions; (iii) to set for moderately high resolution; (iv) to check the stability of the primary ion current frequently; (v) to pay special attention heed to source and collision region pressure stability; and (vi) to keep collision region pressures below 0.6 mPa.