✦ Synopsis


Abstract

When a dissolved oxygen (DO) probe is submerged in an air‐saturated cell culture medium the thickness of the liquid film that exists outside the membrane of a DO probe changes with hydrodynamic shear. The response of the DO probe thus varies with the hydrodynamic shear environment near the DO probe in cell culture reactors. The thickness of the liquid film was estimated by using a three‐layer model, which describes the flow of DO molecules through the liquid layer, the membrane, and the electrolyte, to the cathode of a DO probe. According to the three‐layer model, the current output of the DO probe was a strong function of thickness of the liquid film outside the membrane of the DO probe. A correlation between shear rates on the surface of the probe and the DO saturation reading was obtained by using two concentric cylinders with a rotating inner cylinder. This correlation was then used to characterize the local hydrodynamic shear environment in a cell culture reactor. Β© 1993 John Wiley & Sons, Inc.


πŸ“œ SIMILAR VOLUMES


Determination of dissolved oxygen by use
✍ Antonio Sanz-MartΓ­nez; Angel RΓ­os; Miguel ValcΓ‘rcel πŸ“‚ Article πŸ“… 1993 πŸ› Elsevier Science 🌐 English βš– 406 KB

A simple flow-injection method for the automatic photometric determination of low concentrations of dissolved oxygen in waters was developed. The method uses Methylene Blue immobilized on a cationic exchange resin packed into a flow-cell located in the spectrophotometer, where the leuco form of the

Determination of the volumetric mass tra
✍ L. A. Tribe; C. L. Briens; A. Margaritis πŸ“‚ Article πŸ“… 1995 πŸ› John Wiley and Sons 🌐 English βš– 459 KB πŸ‘ 3 views

There are many dynamic methods for measuring the volumetric mass transfer coefficient. The "gas out-gas in" method can directly determine the volumetric m a s s transfer coefficient in a bioreactor system and provide estimates of the volumetric microbial oxygen uptake rate and the average oxygen sat