## Abstract **Summary:** A material placed in or in contact with a biological system, that causes the minimum perturbation that can be tolerated by the host biological system, can be considered to be biocompatible. The major effects caused by a contact lens can be considered to be 1) the blocking o
Measurement errors related to contact angle analysis of hydrogel and silicone hydrogel contact lenses
✍ Scribed by Michael L. Read; Philip B. Morgan; Carole Maldonado-Codina
- Publisher
- John Wiley and Sons
- Year
- 2009
- Tongue
- English
- Weight
- 132 KB
- Volume
- 91B
- Category
- Article
- ISSN
- 1552-4973
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
This work sought to undertake a comprehensive investigation of the measurement errors associated with contact angle assessment of curved hydrogel contact lens surfaces. The contact angle coefficient of repeatability (COR) associated with three measurement conditions (image analysis COR, intralens COR, and interlens COR) was determined by measuring the contact angles (using both sessile drop and captive bubble methods) for three silicone hydrogel lenses (senofilcon A, balafilcon A, lotrafilcon A) and one conventional hydrogel lens (etafilcon A). Image analysis COR values were about 2°, whereas intralens COR values (95% confidence intervals) ranged from 4.0° (3.3°, 4.7°) (lotrafilcon A, captive bubble) to 10.2° (8.4°, 12.1°) (senofilcon A, sessile drop). Interlens COR values ranged from 4.5° (3.7°, 5.2°) (lotrafilcon A, captive bubble) to 16.5° (13.6°, 19.4°) (senofilcon A, sessile drop). Measurement error associated with image analysis was shown to be small as an absolute measure, although proportionally more significant for lenses with low contact angle. Sessile drop contact angles were typically less repeatable than captive bubble contact angles. For sessile drop measures, repeatability was poorer with the silicone hydrogel lenses when compared with the conventional hydrogel lens; this phenomenon was not observed for the captive bubble method, suggesting that methodological factors related to the sessile drop technique (such as surface dehydration and blotting) may play a role in the increased variability of contact angle measurements observed with silicone hydrogel contact lenses. © 2009 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 2009
📜 SIMILAR VOLUMES
## Abstract __Purpose:__ To evaluate the effect of wear on dynamic __in vitro__ dehydration of silicone hydrogel (Si‐Hi) contact lens (CL) using a previously described gravimetric procedure. __Methods:__ Five different silicone hydrogel (Si‐Hi) contact lenses (CL) were evaluated after being worn by
## Abstract __Purpose:__ The aim of this study was to analyze surface changes of three silicone‐hydrogel contact lenses after daily wear. The lenses used in this study were balafilcon A, lotrafilcon B (both surface‐treated), and galyfilcon A (non surface‐treated). __Methods:__ To understand how and
## Abstract This study investigated the water contact angles of five commercially available silicone hydrogel contact lenses (Acuvue Advance, Acuvue Oasys, Focus Night & Day, O~2~ Optix, and PureVision) using sessile drop and captive bubble techniques. The only lens type that showed a significant d
## Abstract ## Purpose The purpose of the present study was to develop mathematical relationships that allow obtaining equilibrium water content and refractive index of conventional and silicone hydrogel soft contact lenses from refractive index measures obtained with automated refractometry or eq