Three different injection modes were compared for the analysis of organophosphorus pesticides. Differing results were found for several compounds depending on their chemical structure and thermostability. The programmed temperature vaporizing injector used in this study seemed to be the most promisi
Interlaboratory studies of the determination of selected chlorobiphenyl congeners with capillary gas chromatography using splitless- and on-column injection techniques
β Scribed by Tuinstra, L. G. M. Th. ;Roos, A. H. ;Griepink, B. ;Wells, D. E.
- Publisher
- John Wiley and Sons
- Year
- 1985
- Tongue
- English
- Weight
- 580 KB
- Volume
- 8
- Category
- Article
- ISSN
- 0935-6304
No coin nor oath required. For personal study only.
β¦ Synopsis
The Community Bureau of Reference has organized a collaborative interlaboratory project to improve the analytical protocol for some specific chlorobiphenyls (CBs) within the European Community (EC). A series of test procedures were prescribed to optimize the gas chromatographic conditions for splitless and on-column injection, which substantially improved the quality of data. Important parameters included the initial column and injector temperatures and the choice of the polarity of the stationary phase. In a study with cleaned eel-fat extracts, coefficients of variation for reproducibility CV(R) ranging from 11 to 24% at the 0.04-0.3 mg/kg level per CB congener were obtained. No significant difference could be found between splitless-and on-column injection.
π SIMILAR VOLUMES
A specific, sensitive, and precise assay for simultaneous quantitation of glyceryl trinitrate (GTN) and its metabolites 1,2and 1,3-glyceryl dinitrate (GDN) has been developed. The method involves one single extraction step with benzene, followed byevaporation at 55'C.One plof theextractwasinjected o
A specific, sensitive, and accurate assay for quantitation of verapamil has been developed. The method involves a single extraction step with n-heptane, followed by evaporation at room temperature under nitrogen. 0.1 pi of the extract was injected into a capillary column coated with crosslinked 5% p