In vivo laser-induced interstitial thermotherapy of pig liver with a temperature-controlled diode laser and MRI correlation
✍ Scribed by Walter A. Wohlgemuth; Günther Wamser; Thomas Reiss; Theodor Wagner; Klaus Bohndorf
- Publisher
- John Wiley and Sons
- Year
- 2001
- Tongue
- English
- Weight
- 159 KB
- Volume
- 29
- Category
- Article
- ISSN
- 0196-8092
- DOI
- 10.1002/lsm.1131
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Background and Objective
To evaluate the efficacy of a new temperature‐controlled Diode laser to generate in vivo tissue coagulations with laser‐induced interstitial thermotherapy (LITT) in pig liver and MRI‐correlation of these necroses.
Study Design/Materials and Methods
Each of four animal subjects received eight treatments at four different constant temperatures (70, 80, 90, and 100°C) and at two out of three different intervals (3, 6, 9 minutes), resulting in a total of 32 lesions. After sacrificing, the liver was examined in an open MRI (0.35T). Tissue necrosis was measured macroscopically and compared with the size measured by MRI.
Results
The size of the necrosis was energy dependent with the largest extent at 100°C at 9 minutes (2.1 cm). The correlation between macroscopic lesion size and MRI findings was significant (P > 0.036). Histological workup excluded carbonization in every case.
Conclusions
LITT carried out with the Diode laser provides reproducible necroses of pig liver in vivo. MR‐imaging results compared favorably with those obtained by pathologic specimens. Lasers Surg. Med. 29:374–378, 2001. © 2001 Wiley‐Liss, Inc.