## Abstract ## Purpose To evaluate image quality and accuracy for the detection of focal hepatic lesions depicted on T2‐weighted images obtained with two high‐resolution free‐breathing techniques (navigator‐triggered turbo spin‐echo [TSE] and respiratory‐triggered TSE) and two standard‐resolution
Improving the detectability of focal liver lesions on T2-weighted MR images: Ultrafast breath-hold or respiratory-triggered thin-section MRI?
✍ Scribed by Dirk Pauleit; Jochen Textor; Reinald Bachmann; Rudolf Conrad; Sebastian Flacke; Burkhard Kreft; Hans Schild
- Publisher
- John Wiley and Sons
- Year
- 2001
- Tongue
- English
- Weight
- 246 KB
- Volume
- 14
- Category
- Article
- ISSN
- 1053-1807
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
The purpose of this study was to determine whether a respiratory‐triggered (RT) T2‐weighted turbo spin‐echo (TSE) sequence with thin section can improve the detectability of focal liver lesions compared to a breath‐hold (BH) T2‐weighted TSE sequence. In 25 patients an RT TSE with 8‐mm sections (8‐TSE RT) and 5‐mm sections (5‐TSE RT) and a BH TSE sequence with 8‐mm sections (8‐TSE BH) were performed. Forty‐one focal liver lesions (mean: 1.8 ± 1.2 cm; 14 lesions ≤1 cm; 27 lesions >1 cm) were evaluated. The 5‐TSE RT was significantly better in lesion detection compared to the 8‐TSE BH sequence for all sizes of lesions (40/41 vs. 33/41; P = 0.014). For lesions >1 cm no relevant differences in the detection rate of the sequences were found (8‐TSE RT, 26/27; 5‐TSE RT, 26/27; 8‐TSE BH, 25/27), for lesions ≤1 cm the 5‐TSE RT provided significantly better sensitivity than the 8‐TSE BH (14/14 vs. 8/14, P = 0.015). The results of this study suggest that lesion detection could be significantly improved by using an RT TSE sequence with thin sections compared with a BH TSE sequence. J. Magn. Reson. Imaging 2001;14:128–133. © 2001 Wiley‐Liss, Inc.
📜 SIMILAR VOLUMES