Improved Čerenkov counting of 32P for liquid scintillation counter
✍ Scribed by Paul N.P. Chow
- Publisher
- Elsevier Science
- Year
- 1980
- Weight
- 111 KB
- Volume
- 31
- Category
- Article
- ISSN
- 0020-708X
No coin nor oath required. For personal study only.
📜 SIMILAR VOLUMES
## Application to Photophosphorylation Gerenkov radiation produced by hard p-emitters has been applied to the assay of radioisotopes in a liquid scintillation counter without the addition of scintillators (1,2). In aqueous samples, the P-rays having energies greater than 0.263 MeV are expected to
The effect of sample vial type and sample composition on the Cerenkov count rate detected from 32P and 36Cl was studied using a liquid scintillation counter. When counting was done in the noncoincident mode, glass vials allowed higher counting efficiency than plastic vials. In the coincident mode li