## Abstract The __T__~2~\* relaxation time of blood varies with its oxygen saturation. To evaluate the feasibility of imaging intravascular blood oxygenation in humans using a conventional 1.5T MR system, we have implemented a method to measure __T__~2~\* of blood despite the presence of pulsatile
High-speed black blood imaging of vessel stenosis in the presence of pulsatile flow
✍ Scribed by Daisy Chien; Alexander Goldmann; Robert R. Edelman
- Publisher
- John Wiley and Sons
- Year
- 1992
- Tongue
- English
- Weight
- 518 KB
- Volume
- 2
- Category
- Article
- ISSN
- 1053-1807
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Stenosis phantoms were created to study the ability of “black blood” methods to image a vessel stenosis in the presence of pulsatile flow. Black blood images were acquired with a modified TurboFLASH (fast low‐angle shot) method that eliminates flow signal by applying a set of prepulses before segmented data acquisition. With this high‐speed approach, imaging can be completed within 16 seconds. This technique was compared with conventional spin‐echo black blood, gradient‐echo black blood, and gradient‐echo bright blood methods. Loss of flow signal, which extended beyond the site of the stenosis, was seen on the gradient‐echo bright blood images. The pattern of signal loss varied with the type of stenosis. Flow voids were achieved with spin‐echo black blood imaging: however, substantial ghosting artifacts were seen. With gradient‐echo black blood imaging, it was difficult to eliminate all flow signal, particularly for in‐plane flow. The modified TurboFLASH method produced high‐quality black blood images in a fraction of the time needed for spin‐echo imaging. It showed no ghosting artifacts even in the presence of pulsatile flow.
📜 SIMILAR VOLUMES
## Abstract ## Purpose: To measure uterine artery and vein blood velocity and flow rate profiles using MRI during normal pregnancy. ## Materials and Methods: A two‐shot velocity magnitude‐encoded echo planar imaging (EPI) sequence is used at a magnetic field 0.5T. Data analysis procedures, neces