Glass wool in the inserts of split injectors for capillary GC?

✍ Scribed by Grob, K. ;Neukom, H. P. ;Hilling, P.


Publisher
John Wiley and Sons
Year
1981
Tongue
English
Weight
593 KB
Volume
4
Category
Article
ISSN
0935-6304

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

It has been reported that glass wool packed tightly into the glass liner of a vaporizing injector used in the splitting mode considerably reduces the standard deviation of the results obtained, because of improved evaporation of the sample prior to reaching the split point at the capillary column entrance. This finding could not be reproduced on using the same sample composition as reported in the literature, i.e. methanol/2‐ethyl‐1‐hexanol (1:1).

The standard deviations obtained were between 3 and 10% (depending on the conditions selected) and were not influenced significantly by the introduction of glass wool.

The peak area ratio (methanol/2‐ethyl‐1‐hexanol) was found to depend on a number of parameters, such as: injector temperature; glass liner internal diameter; syringe handling technique; the relative position of the syringe needle exit and capillary column entrance; the sample volume injected; and the packing of the glass liner. Generally, the area ratio deviated further from the correct one (determined by cold on‐column sampling) when the glass liner was packed with glass wool.

On the basis of the results, it is speculated that either a complete evaporation of the sample should be achieved (which appears to be impossible under the conditions we used) or, alternatively, the sample should be given the least possible opportunity to evaporate, thus allowing it to enter the column in the form of droplets. The results were worse in terms of precision and accuracy the greater the partition of sample components between the liquid (droplet) and vapor phase. It is concluded that the use of evaporation aids such as glass wool cannot be generally recommended.


πŸ“œ SIMILAR VOLUMES


Sample evaporation in conventional split
✍ Grob, Konrad ;De Martin, Mirella πŸ“‚ Article πŸ“… 1992 πŸ› John Wiley and Sons 🌐 English βš– 753 KB

## Abstract Perylene is strongly fluorescent as long as it is in solution. This has enabled visual observation of non‐evaporated sample material in a β€œtransparent injector”, __i.e.__ in a heated glass device imitating a conventional vaporizing injector. Three scenarios of sample evaporation are de

The influence of evaporative dynamics on
✍ Bowermaster, J. πŸ“‚ Article πŸ“… 1988 πŸ› John Wiley and Sons 🌐 English βš– 573 KB

The kineticsof sample evaporation wasstudied for four common injection liners at various temperatures.The ratesof solvent and sample evaporation were measured. The sample distribution at the split point was probed by inserting two capillary columns in one injector. Greater homogeneity at the split p

Potential and limitations of differently
✍ Staniewski, Jacek ;Rijks, Jacques A. πŸ“‚ Article πŸ“… 1993 πŸ› John Wiley and Sons 🌐 English βš– 617 KB

## Abstract Programmed temperature sample introduction is an attractive technique both for the introduction of large volumes of dilute samples in capillary gas chromatography and for LC‐GC interfacing. In this study a new type of liner containing a porous bed of sintered glass is introduced and ev