## Abstract ## Purpose The purpose of this study is to evaluate the feasibility and image quality of multishot T2‐weighted (T2w) renal morphologic imaging based on the acquisition of rotating rectangular strips of k‐space data after successive radiofrequency excitations (BLADE technique). ## Mate
Fully balanced steady-state 3D-spin-echo (bSSSE) imaging at 3 Tesla
✍ Scribed by R. Krug; E.T. Han; S. Banerjee; S. Majumdar
- Book ID
- 102952868
- Publisher
- John Wiley and Sons
- Year
- 2006
- Tongue
- English
- Weight
- 582 KB
- Volume
- 56
- Category
- Article
- ISSN
- 0740-3194
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
In this work a 3D fully balanced steady‐state spin‐echo sequence (bSSSE) was designed and implemented on a 3‐Tesla whole‐body MR scanner. In contrast to previously introduced nonbalanced SSSE (nbSSSE) sequences, bSSSE features fully rewound imaging gradients. Additionally, all phase encodings and the readout prephasing are performed after the refocusing pulse. Its performance was compared with previously used gradient‐echo (GE) and nbSSSE sequences. The signal response of each sequence was simulated by numerical solution of the Bloch equation. All sequences were then optimized based on these simulations for high‐resolution (HR) imaging of tissues with short relaxation times and applied to in vivo HR‐MRI of trabecular bone. bSSSE outperformed nbSSSE sequences in terms of SNR and SNR efficiency. Measurements of structural bone parameters revealed thinner trabeculae and a lower bone/bone‐marrow fraction for the bSSSE sequence compared to the nbSSSE sequence. By applying a parallel imaging technique based on generalized autocalibrating partially parallel acquisition (GRAPPA) with a reduction factor of 2, we were able to maintain a clinically feasible scan time using bSSSE. Comparisons of structural bone parameters revealed a difference of less than 3% for all structural parameters between parallel and conventional imaging. Magn Reson Med, 2006. © 2006 Wiley‐Liss, Inc.
📜 SIMILAR VOLUMES
## Abstract Whole‐brain echo‐planar spectroscopic imaging (EPSI) often substantially lengthens MRI/MRSI (magnetic resonance spectroscopic imaging) protocols. To halve acquisition time, application of a blipped phase‐encoding (PE) gradient during the EPSI readout (RO) was previously suggested by PE
## Abstract The double‐echo‐steady‐state (DESS) sequence generates two signal echoes that are characterized by a different contrast behavior. Based on these two contrasts, the underlying T2 can be calculated. For a flip‐angle of 90°, the calculated T2 becomes independent of T1, but with very low si