## Abstract ## Purpose: To evaluate motion artifacts, tissue contrasts, and lesion detectability in the neck with the periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER) (BLADE) technique. ## Materials and Methods: A total of 46 patients referred for MRI of t
Renal T2-weighted turbo-spin-echo imaging with BLADE at 3.0 Tesla: Initial experience
✍ Scribed by Henrik J. Michaely; Harald Kramer; Sabine Weckbach; Olaf Dietrich; Maximilian F. Reiser; Stefan O. Schoenberg
- Publisher
- John Wiley and Sons
- Year
- 2007
- Tongue
- English
- Weight
- 200 KB
- Volume
- 27
- Category
- Article
- ISSN
- 1053-1807
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Purpose
The purpose of this study is to evaluate the feasibility and image quality of multishot T2‐weighted (T2w) renal morphologic imaging based on the acquisition of rotating rectangular strips of k‐space data after successive radiofrequency excitations (BLADE technique).
Materials and Methods
A total of 7 healthy volunteers and 27 patients with suspected renal and renovascular diseases were included in this prospective intraindividual study. All exams were performed at 3.0T in a random order with a standard T2w turbo‐spin‐echo (TSE) sequence and with a T2w‐BLADE sequence with equal spatial resolution. Phantom measurements were performed to measure the objective signal‐to‐noise ratio (SNR). Two radiologists rated the image sharpness, the flow signal suppression, the presence of artifacts, and the overall image quality of both techniques and determined their preferred sequence.
Results
The SNR did not show significant differences. The overall image quality was rated significantly higher for the T2w‐BLADE (P < 0.05). Equally, the presence of disturbing artifacts and the sharpness was ranked significantly better for T2w‐BLADE than for the T2w‐TSE (P < 0.05). The T2w‐BLADE was the preferred sequence in 59% of all cases, the T2w‐TSE in 9%.T2w‐BLADE sequences seem to be superior for the depiction of the kidneys at 3.0T, particularly due to the decreased number of artifacts and sharper delineation of the organs. J. Magn. Reson. Imaging 2007. © 2007 Wiley‐Liss, Inc.