Flow artifacts in steady-state free precession cine imaging

✍ Scribed by Pippa Storey; Wei Li; Qun Chen; Robert R. Edelman


Publisher
John Wiley and Sons
Year
2003
Tongue
English
Weight
300 KB
Volume
51
Category
Article
ISSN
0740-3194

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

Steady‐state free precession (SSFP) cardiac cine images are frequently corrupted by dark flow artifacts, which can usually be eliminated by reshimming and retuning the scanner. A theoretical explanation for these artifacts is provided in terms of spins moving through an off‐resonant point in the magnetic field, and the theory is validated using phantom experiments. The artifacts can be reproduced in vivo by detuning the center frequency by an amount in the range of half the inverse repetition time (TR). Since this offset is similar in magnitude to the frequency difference between the water and lipid peaks, a likely cause of the artifacts in vivo is that the center frequency is tuned incorrectly to the lipid peak rather than the water peak. Magn Reson Med 51:115–122, 2004. © 2003 Wiley‐Liss, Inc.


📜 SIMILAR VOLUMES


Flow effects in balanced steady state fr
✍ M. Markl; M.T. Alley; C.J. Elkins; N.J. Pelc 📂 Article 📅 2003 🏛 John Wiley and Sons 🌐 English ⚖ 876 KB

## Abstract An analysis of the effect of flow on 2D fully balanced steady state free precession (SSFP) imaging is presented. Transient and steady‐state SSFP signal intensities in the presence of steady and pulsatile flow were simulated using a matrix formalism based on the Bloch equations. Various

Frequency-modulated steady-state free pr
✍ D.L. Foxall 📂 Article 📅 2002 🏛 John Wiley and Sons 🌐 English ⚖ 800 KB

Exploration of the possibilities of steady-state free precession (SSFP) excitation has led to the discovery that it is tolerant of slow variations in spectral offset frequency. The effect has been used to eliminate banding artifacts from images obtained with the fully balanced SSFP imaging sequence.

Motion-insensitive, steady-state free pr
✍ Y. Zur; M. L. Wood; L. J. Neuringer 📂 Article 📅 1990 🏛 John Wiley and Sons 🌐 English ⚖ 788 KB

## Abstract Steady‐state free precession (SSFP) pulse sequences employing gradient reversal echoes and short repetition time (TR) between successive rf excitation pulses offer high signal‐to‐noise ratio per unit time. However, SSFP sequences are very sensitive to motion. A new SSFP method is presen

Steady state free precession magnetizati
✍ Oliver Bieri; Tallas C. Mamisch; Siegfried Trattnig; Klaus Scheffler 📂 Article 📅 2008 🏛 John Wiley and Sons 🌐 English ⚖ 647 KB

## Abstract The formerly proposed concept for magnetization transfer imaging (MTI) using balanced steady‐state free precession (SSFP) image acquisitions is in this work extended to nonbalanced protocols. This allows SSFP‐based MTI of targets with high susceptibility variation (such as the musculosk

Cardiac CINE imaging with IDEAL water-fa
✍ Scott B. Reeder; Michael Markl; Huanzhou Yu; Jeffrey C. Hellinger; Robert J. Her 📂 Article 📅 2005 🏛 John Wiley and Sons 🌐 English ⚖ 889 KB

## Abstract ## Purpose to decompose multicoil CINE steady‐state free precession (SSFP) cardiac images acquired at short echo time (TE) increments into separate water and fat images, using an iterative least‐squares “Dixon” (IDEAL) method. ## Materials and Methods Multicoil CINE IDEAL‐SSFP cardia

Quantitative diffusion imaging with stea
✍ Sean C.L. Deoni; Terry M. Peters; Brian K. Rutt 📂 Article 📅 2004 🏛 John Wiley and Sons 🌐 English ⚖ 154 KB

## Abstract The addition of a single, unbalanced diffusion gradient to the steady‐state free precession (SSFP) imaging sequence sensitizes the resulting signal to free diffusion. Unfortunately, the confounding influence of both longitudinal (__T__~__1__~) and transverse (__T__~__2__~) relaxation on