Field evaluation of precision of three capillary blood glucose monitors
β Scribed by Dougherty, RC ;Niu, X-F ;Deeb, LC
- Publisher
- John Wiley and Sons
- Year
- 1994
- Tongue
- English
- Weight
- 301 KB
- Volume
- 11
- Category
- Article
- ISSN
- 1357-8170
No coin nor oath required. For personal study only.
β¦ Synopsis
Abstract
This paper discusses a procedure for independent evaluation of commercially available blood glucose meters. The evaluation exercise provides a basis for assessment of manufacturer's claims. It also provides solid experience in the reliability of blood glucose data obtained from a specific meter under field conditions.
We report measurements of the precision of three capillary blood glucose measurement devices, commercially available in the USA, used under field conditions. We have compared the measurements of blood glucose samples analysed by three different instruments. The method of analysis involves making two blood glucose measurements from a single finger puncture. This experimental design in available to any clinician who would wish to try the expeiriment. We report values for the precision (reproducibility) of a single device and comparisons for pairs of meters using standard statistical techniques. The coefficient of variation for the difference between the first and second measurements obtained for three instruments was: A, Β±28.7%; B, Β±36.5%; C, Β±10.2%. The data show that instrument C worked much better than either A or B for reproducing blood glucose values on the same sample under field conditions.
π SIMILAR VOLUMES
## Abstract Capillary blood glucose monitoring plays an important part in the monitoring of diabetes. There are a large number of different blood glucose meters, each of which have satisfactory analytical performance for monitoring purposes in the hands of experienced personnel. Of less certainty i
The superficial capillary network of the gastric mucosa can be monitored for red blood cell velocity measurements by a microscopic technique. This network, however, reflects the blood flow in capillaries of more physiological interest, namely those passing by the acid-producing cells and emptying in
concentration in 48 patients, who were breathing high concentrations of oxygen, was investigated. Arterial and pulmonary arterial blood glucose values were analyzed using the GOD-F method and, as a control, the hexokinase method. The respective pO 2 values were also measured. The blood glucose conce