USA - Comprehensive evaluation of blood glucose self-monitoring system : In CURR. THER. RES. CLIN. EXP. (55/10 (1127–1135) 1994) J. R. White. Jr, R. K. Campbell, A. Freerksen, B. Gould of Washington State University report on “A comprehensive evaluation of a blood glucose self-monitoring system for diabetes care”
✍ Scribed by Valerie M. Owen
- Publisher
- Elsevier Science
- Year
- 1995
- Tongue
- English
- Weight
- 189 KB
- Volume
- 10
- Category
- Article
- ISSN
- 0956-5663
No coin nor oath required. For personal study only.
✦ Synopsis
concentration in 48 patients, who were breathing high concentrations of oxygen, was investigated. Arterial and pulmonary arterial blood glucose values were analyzed using the GOD-F method and, as a control, the hexokinase method. The respective pO 2 values were also measured. The blood glucose concentrations measured by the GOD-F method show a significant linear relation with that measured by the hexokinase method in both arterial (y = -24.4 + 1.01x, r = 0.99) and pulmonary arterial blood (y = -3.4 + 1.01x, r = 0.96). The difference of intercepts is statistically significant, but the error of the GOD-F method could not be assessed just by the difference. The equation defining the effect ofpO2 on the percent change between blood glucose measured by the GOD-F method and that measured by the hexokinase method generally follows our measured materials and introduces the relationship among blood glucose value, pO2, and the error of the GOD-F method.