𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Fat-suppressed 3D-T1-weighted-echo planar imaging: comparison with fat-suppressed 3D-T1-weighted-gradient echo in imaging the cartilage of the knee

✍ Scribed by Apostolos H. Karantanas; Aristidis H. Zibis; Panagiotis Kitsoulis


Book ID
104350426
Publisher
Elsevier Science
Year
2002
Tongue
English
Weight
299 KB
Volume
26
Category
Article
ISSN
0895-6111

No coin nor oath required. For personal study only.

✦ Synopsis


This study was conducted to compare a three-dimensional (3D) multi-shot echo-planar imaging (EPI) sequence with fat-suppression (FS) with the 3D-fat-suppressed gradient echo (GRE-FS) sequence in imaging the cartilage of the knee. One hundred sixty-nine patients were studied prospectively. The cartilage was imaged in the sagittal plane with: (a) 3D-T1-EPI-FS and (b) 3D-T1-GRE-FS sequences using a 1 T MR scanner. The signal-to-noise ratio (SNR) of bone (b) and cartilage (c), and relative contrast (ReCon) between bone and cartilage and meniscus and cartilage were measured in 60 patients with arthroscopically normal cartilage. The imaging accuracy was assessed by comparing with linear regression analysis (length and depth) 32 defects in the cartilage of cadaveric (human and bovine) knees. The 3D-T1-EPI-FS provided better bone marrow signal suppression, better SNRc and better ReCon(bc) and ReCon(cm) (p , 0.01). The 3D-T1-EPI-FS showed better accuracy concerning the depth of the defects and the 3D-T1-GRE-FS better accuracy concerning the length of the defects. In conclusion, the 3D-T1-EPI-FS pulse sequence could be included in the routine protocol in imaging the cartilage of the knee because it achieves high SNR of the cartilage and high ReCon compared to the surrounding structures, at a reduced scan time.


📜 SIMILAR VOLUMES


Reversed laminar appearance of articular
✍ Yang Xia; Shaokuan Zheng 📂 Article 📅 2010 🏛 John Wiley and Sons 🌐 English ⚖ 443 KB

## Abstract ## Purpose: To investigate the reversed intensity pattern in the laminar appearance of articular cartilage by 3D fat‐suppressed spoiled gradient recalled echo (FS‐SPGR) imaging in magnetic resonance imaging (MRI). ## Materials and Methods: The 3D SPGR experiments were carried out on

Imaging of the articular cartilage in os
✍ Hiroshi Yoshioka; Marcus Alley; Daniel Steines; Kathryn Stevens; Erika Rubesova; 📂 Article 📅 2003 🏛 John Wiley and Sons 🌐 English ⚖ 435 KB

## Abstract ## Purpose To compare three‐dimensional (3D) spatial‐spectral (SS) spoiled gradient‐recalled acquisition in the steady state (SPGR) imaging with fat‐suppressed 3D SPGR sequences in MR imaging of articular cartilage of the knee joint in patients with osteoarthritis. ## Materials and Me

T2-weighted 3D fast spin echo imaging wi
✍ Ananth J. Madhuranthakam; Huanzhou Yu; Ann Shimakawa; Reed F. Busse; Martin P. S 📂 Article 📅 2010 🏛 John Wiley and Sons 🌐 English ⚖ 375 KB

## Abstract ## Purpose: To develop a robust 3D fast spin echo (FSE) T~2~‐weighted imaging method with uniform water and fat separation in a single acquisition, amenable to high‐quality multiplanar reformations. ## Materials and Methods: The Iterative Decomposition of water and fat with Echo Asym