## Ab!stract This paper describes the evaluation of several data-processmg optIons for a flow-based sample processor with a well-stIrred nuxmg chamber Data-processing options evaluated Include peak-height, peak-area and tune-mterval methods v&h dtierent chorces of reference pomts Results are compa
โฆ LIBER โฆ
Experimental evaluation of theoretical response equations for an unsegmented flow system with a well-stirred mixing chamber
โ Scribed by James M. Jordan; Harry L. Pardue
- Publisher
- Elsevier Science
- Year
- 1992
- Tongue
- English
- Weight
- 699 KB
- Volume
- 270
- Category
- Article
- ISSN
- 0003-2670
No coin nor oath required. For personal study only.
โฆ Synopsis
This paper describes results of an experimental study of theoretical equations developed to describe time-dependent responses of a flow-based sample-processing system that includes a well-stirred mixing chamber. It also describes an alternative approach to the sample-introduction step that improves agreement between experimental and theoretical results. The study confirms the validity of the theoretical equations for situations in which the assumptions on which they are based are met. The most serious deviations occur for situations in which chemical reactions are slow.