## Abstract Three breast tissue expanders were evaluated for compatibility with MR imaging (1.5 Tesla). The metallic components of the breast tissue expanders were shown to be nonferromagnetic, heating .2Β°C and the artifacts varied. These results indicate that MR procedures may be performed safely
Ex vivo evaluation of ferromagnetism and artifacts of cardiac occluders exposed to a 1.5-T MR system
β Scribed by Frank G. Shellock; Stacy M. Morisoli
- Publisher
- John Wiley and Sons
- Year
- 1994
- Tongue
- English
- Weight
- 360 KB
- Volume
- 4
- Category
- Article
- ISSN
- 1053-1807
No coin nor oath required. For personal study only.
β¦ Synopsis
Abstract
Magnetic resonance (MR) procedures are contraindicated for patients with certain ferromagnetic biomedical implants, primarily owing to the risk of movement or dislodgment of the implants by the static magnetic field. An additional concern is the amount of artifact that the implant produces, which can affect image quality and interpretation of the examination. Therefore, an ex vivo assessment of ferromagnetism and artifact was conducted for 12 different occluders used to treat patients with patent ductus arteriosus, atrial septal defects, and ventricular septal defects, in a 1.5βT MR system. Seven of the occluders, made of 304 stainless steel, were ferromagnetic and displayed deflection forces of 248β299 dynes. Five of the implants, made of MP35n, were nonferromagnetic. Artifacts were variable and depended primarily on the type and amount of metal used to construct the implant. The authors conclude that patients with ferromagnetic cardiac occluders may undergo MR procedures approximately 6 weeks after placement of these devices, to allow tissue growth to provide additional retentive force. After this time, it is highly unlikely that the magnetic fields associated with a 1.5βT MR system are capable of moving or dislodging any of these cardiac occluders.
π SIMILAR VOLUMES
## Abstract Ex vivo testing techniques were used to determine the ferromagnetic qualities of, presence of heating in, and artifacts produced by 13 different heart valve prostheses exposed to a 1.5βT (64βMHz) magnetic resonance (MR) system. None of the heart valve prostheses showed a measurable defl
## Abstract The purpose of our study was to asacsa Futomagneh. hating. ad artifactr, associated with vabcular access ports exposed to a 1.5βT MR system. Tweatyβcight different vabcular access port. were evaluated in thb investlgation. Fmmagnetism was determined by using two previouely de techniques
## Forty -eight biopsy devices (needles, stylets, guide wires. biopsy guns. etc) were evaluated for ferromagnetism (n = 48) and artifacts (n = 43) in a 1.5-T magnetic resonance (MR) system, with previously deflve of the 48 devices were found to be ferromagnetic. Artifacts varied, depending on the