screening concepts for diabetic nephropathy. Part of this project was the evaluation of the newly developed qualitative immunologic self-test Micral-Test ® S.
Evaluation of the micral-test® as a screening procedure for micro-albuminuria
✍ Scribed by J. Massé; M. Pouliot; P. Leclerc; B. Fruteau
- Publisher
- Elsevier Science
- Year
- 1995
- Tongue
- English
- Weight
- 123 KB
- Volume
- 28
- Category
- Article
- ISSN
- 0009-9120
No coin nor oath required. For personal study only.
✦ Synopsis
Reml~ At a cut-off of 20 mg/L, the Ivfioral-Test showed a poor senmtivity (56%). Correction for creafininc did incresee the sensitivity (69%) for the same specificity (98.7"A). The beat screening performance was obtained when the I~emI-Te=t results were expressed as rate 0tg/min) with a recciver-o~ttor characteristic (ROC) curve area of 0.96. Conclusion Semi-qusatitativ,; determination of mieroalbuminuda on an ovemi#t tpecimen shows an inadequate sensitivity unless it is comoted for udue conecntrstion (ereetinine). A better serecning performance is obtained by using a timed specimen and calculatinn of micrcalbumin excretion rote.
📜 SIMILAR VOLUMES
At this present time, we feel that there is no role for DNA flow cytometry (FCM), or indeed DNA studies by any other method, to be used as a screening procedure for patients with no prior history of bladder cancer due to the high false-positive rate found when monitoring exfoliated urothelial cells.
To assess the performance of different cutoff points for fasting plasma glucose (FPG) including the World Health Organization (WHO) cutoff of FPG у7.8 mmol l -1 for the screening of diabetes mellitus, data from oral glucose tolerance tests done on 865 Singaporean adults over a 4-year period were ana