Evaluation of respiratory liver and kidney movements for MRI navigator gating

✍ Scribed by Ruitian Song; Aaryani Tipirneni; Perry Johnson; Ralf B. Loeffler; Claudia M. Hillenbrand


Book ID
102379463
Publisher
John Wiley and Sons
Year
2010
Tongue
English
Weight
382 KB
Volume
33
Category
Article
ISSN
1053-1807

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

Purpose

To determine the tracking factor by studying the relationship between kidney and diaphragm motions and to compare the efficiency of the gating‐and‐following and gating‐only algorithms in reducing motion artifacts in navigator‐gated scans.

Materials and Methods

Diaphragm and kidney motions were measured by using real‐time TrueFISP sequences from 10 healthy human volunteers to determine tracking factors at different acceptance windows. Mean tracking factors were used to calculate mean residual errors and improvement factors for the gating‐and‐following and gating‐only algorithms.

Results

Mean tracking factors for ±4, ±6, ±8 mm and full acceptance windows ranged from 0.6 to 0.7, with large interindividual variations. Acceptance rates increased as the size of the acceptance window increased (acceptance rate for a 4 mm window ∼50%). There was a greater reduction of motion errors by gating‐and‐following (maximum of 1.86 mm) than gating‐only (maximum of 7.05 mm).

Conclusion

Mean tracking factors obtained in this study can be used as a guideline for using the gating‐and‐following algorithm in navigator‐gated kidney scans. The gating‐and‐following and gating‐only algorithms were quantitatively compared, and it was found that the former is more effective in reducing motion errors. J. Magn. Reson. Imaging 2011;33:143–148. © 2010 Wiley‐Liss, Inc.


📜 SIMILAR VOLUMES


Preliminary investigation of respiratory
✍ Andrew C. Larson; Peter Kellman; Andrew Arai; Glenn A. Hirsch; Elliot McVeigh; D 📂 Article 📅 2004 🏛 John Wiley and Sons 🌐 English ⚖ 365 KB

## Abstract Segmented cine MRI generally requires breath‐holding, which can be problematic for many patients. Navigator echo techniques, particularly successful for free‐breathing coronary MRA, are incompatible with the acquisition strategies and SSFP pulse sequences commonly used for cine MRI. The

Navigator-triggered oxygen-enhanced MRI
✍ Francesco Molinari; Monika Eichinger; Frank Risse; Christian Plathow; Michael Pu 📂 Article 📅 2007 🏛 John Wiley and Sons 🌐 English ⚖ 372 KB

## Abstract ## Purpose To evaluate an optimized method for oxygen‐enhanced MRI of the lung, using simultaneous electrocardiograph (ECG) and navigator triggering. To correlate oxygen‐enhanced MRI with lung function tests assessing alveolar‐capillary gas exchange. ## Materials and Methods A total