๐”– Bobbio Scriptorium
โœฆ   LIBER   โœฆ

ET-AAS Determination of Trace Analytes in High Purity Bismuth and Tellurium Oxides

โœ Scribed by I. Karadjova


Publisher
Elsevier Science
Year
1996
Tongue
English
Weight
202 KB
Volume
54
Category
Article
ISSN
0026-265X

No coin nor oath required. For personal study only.

โœฆ Synopsis


Optimal conditions for direct electrothermal atomic absorption (ET-AAS) determination of Al, Cd, Co, Cr, Cu, Fe, Mn, Ni, Pb, and V in high purity bismuth oxide and tellurium oxide are described. Matrix interference on the atomization of the analytes is studied. The analytical method developed in this paper permits determination of 0.005 mg g 01 Cd and Mn; 0.02-0.05 mg g 01 Al, Cu, Cr, and Fe; 0.1 mg g 01 Co, Ni, and Pb; 0.5 mg g 01 V with relative standard deviation 3-8%.