𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Ensuring safety of implanted devices under MRI using reversed RF polarization

✍ Scribed by William R. Overall; John M. Pauly; Pascal P. Stang; Greig C. Scott


Publisher
John Wiley and Sons
Year
2010
Tongue
English
Weight
528 KB
Volume
64
Category
Article
ISSN
0740-3194

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

Patients with long‐wire medical implants are currently prevented from undergoing magnetic resonance imaging (MRI) scans due to the risk of radio frequency (RF) heating. We have developed a simple technique for determining the heating potential for these implants using reversed radio frequency (RF) polarization. This technique could be used on a patient‐to‐patient basis as a part of the standard prescan procedure to ensure that the subject's device does not pose a heating risk. By using reversed quadrature polarization, the MR scan can be sensitized exclusively to the potentially dangerous currents in the device. Here, we derive the physical principles governing the technique and explore the primary sources of inaccuracy. These principles are verified through finite‐difference simulations and through phantom scans of implant leads. These studies demonstrate the potential of the technique for sensitively detecting potentially dangerous coupling conditions before they can do any harm. Magn Reson Med, 2010. © 2010 Wiley‐Liss, Inc.


📜 SIMILAR VOLUMES


Measuring RF-induced currents inside imp
✍ Peter Nordbeck; Ingo Weiss; Philipp Ehses; Oliver Ritter; Marcus Warmuth; Floria 📂 Article 📅 2009 🏛 John Wiley and Sons 🌐 English ⚖ 684 KB

## Abstract Radiofrequency (RF)‐related heating of cardiac pacemaker leads is a serious concern in magnetic resonance imaging (MRI). Recent investigations suggest such heating to be strongly dependent on an implant's position within the surrounding medium, but this issue is currently poorly underst