Measuring RF-induced currents inside implants: Impact of device configuration on MRI safety of cardiac pacemaker leads

✍ Scribed by Peter Nordbeck; Ingo Weiss; Philipp Ehses; Oliver Ritter; Marcus Warmuth; Florian Fidler; Volker Herold; Peter M. Jakob; Mark E. Ladd; Harald H. Quick; Wolfgang R. Bauer


Publisher
John Wiley and Sons
Year
2009
Tongue
English
Weight
684 KB
Volume
61
Category
Article
ISSN
0740-3194

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

Radiofrequency (RF)‐related heating of cardiac pacemaker leads is a serious concern in magnetic resonance imaging (MRI). Recent investigations suggest such heating to be strongly dependent on an implant's position within the surrounding medium, but this issue is currently poorly understood. In this study, phantom measurements of the RF‐induced electric currents inside a pacemaker lead were performed to investigate the impact of the device position and lead configuration on the amount of MRI‐related heating at the lead tip. Seven hundred twenty device position/lead path configurations were investigated. The results show that certain configurations are associated with a highly increased risk to develop MRI‐induced heating, whereas various configurations do not show any significant heating. It was possible to precisely infer implant heating on the basis of current intensity values measured inside a pacemaker lead. Device position and lead configuration relative to the surrounding medium are crucial to the amount of RF‐induced heating in MRI. This indicates that a considerable number of implanted devices may incidentally not develop severe heating in MRI because of their specific configuration in the body. Small variations in configuration can, however, strongly increase the risk for such heating effects, meaning that hazardous situations might appear during MRI. Magn Reson Med, 2009. © 2009 Wiley‐Liss, Inc.


📜 SIMILAR VOLUMES


Impact of imaging landmark on the risk o
✍ Peter Nordbeck; Oliver Ritter; Ingo Weiss; Marcus Warmuth; Daniel Gensler; Natal 📂 Article 📅 2010 🏛 John Wiley and Sons 🌐 English ⚖ 849 KB

## Abstract Implanted medical devices such as cardiac pacemakers pose a potential hazard in magnetic resonance imaging. Electromagnetic fields have been shown to cause severe radio frequency‐induced tissue heating in some cases. Imaging exclusion zones have been proposed as an instrument to reduce