Energy resolution of a photon-counting silicon strip detector
✍ Scribed by Erik Fredenberg; Mats Lundqvist; Björn Cederström; Magnus Åslund; Mats Danielsson
- Publisher
- Elsevier Science
- Year
- 2010
- Tongue
- English
- Weight
- 457 KB
- Volume
- 613
- Category
- Article
- ISSN
- 0168-9002
No coin nor oath required. For personal study only.
✦ Synopsis
A photon-counting silicon strip detector with two energy thresholds was investigated for spectral X-ray imaging in a mammography system. Preliminary studies already indicate clinical benefit of the detector, and the purpose of the present study is optimization with respect to energy resolution. Factors relevant for the energy response were measured, simulated, or gathered from previous studies, and used as input parameters to a cascaded detector model. Threshold scans over several X-ray spectra were used to calibrate threshold levels to energy, and to validate the model. The energy resolution of the detector assembly was assessed to range over DE=E ¼ 0:1220:26 in the mammography region.
Electronic noise dominated the peak broadening, followed by charge sharing between adjacent detector strips, and a channel-to-channel threshold spread. The energy resolution may be improved substantially if these effects are reduced to a minimum. Anti-coincidence logic mitigated double counting from charge sharing, but erased the energy resolution of all detected events, and optimization of the logic is desirable. Pile-up was found to be of minor importance at typical mammography rates.
📜 SIMILAR VOLUMES
Double-sided silicon strip detectors (DSSD) with an energy resolution of 1-2 keV full-width at half-maximum (FWHM) are attractive devices for future hard X-ray and soft g-ray applications. For example, they are well suited as scatterer detectors for semiconductor Compton telescopes working in the su
In this paper, we present the design and preliminary performance evaluation of a novel energy-resolved photon-counting (ERPC) detector for gamma ray imaging applications. The prototype ERPC detector has an active area of 4.4 cm × 4.4 cm, which is pixelated into 128 × 128 square pixels with a pitch s