Atomization characteristics of chromium have been studied by electrothermal atomic absorption spectrometry with a molybdenum tube atomizer. The appearance temperature (T app ) of chromium nitrate was 1300ΠC. The T app was independent of the heating rate of atomizer, but the temperature at the peak o
Electrothermal Atomic Absorption Spectrometric Determination of Potassium in Biological Materials with a Molybdenum Tube Atomizer
β Scribed by K. Ohta; K. Mizutani; S. Itoh; T. Mizuno
- Publisher
- Elsevier Science
- Year
- 1993
- Tongue
- English
- Weight
- 388 KB
- Volume
- 48
- Category
- Article
- ISSN
- 0026-265X
No coin nor oath required. For personal study only.
β¦ Synopsis
Electrothermal atomic absorption spectrometry (ETA-AAS) of potassium in biological materials with a molybdenum tube atomizer has been investigated. The atomization characteristics of potassium on the atomizer are demonstrated. The optimal gas flow rate was (400 \mathrm{ml} \mathrm{min}^{-1} \mathrm{Ar}) and (100 \mathrm{ml} \mathrm{min}^{-1} \mathrm{H}_{2}). The characteristic mass (the mass of element giving 0.0044 abs.) of potassium by the atomizer was (75 \mathrm{fg}), which is better than that obtained with a graphite atomizer. The detection limit ((s / n=3)) was (72 \mathrm{fg}) (corresponding to (7.1 \mathrm{pg} \mathrm{ml}^{-1}), with (10 \mu) given). The value was also better than those from flame AA, graphite furnace AA, and flame atomic emission. The relative standard deviation by the use of the molybdenum atomizer was (2.6 %). Interferences caused by large amounts of interferents were evaluated. Thiourea as the matrix modifier eliminated the interferences. The determined potassium in biological materials was in good agreement with the certified values. For a graphite furnace AAS, the recommended method will serve as a highly sensitive determination of potassium in biological materials, when a molybdenum or tungsten metal foil lined into the graphite furnace is used in argon + hydrogen purge gas. 1993 Academic Press. Inc.
π SIMILAR VOLUMES
Ultrasonic slurry sampling electrothermal atomic absorption spectrometry with a metal tube atomizer has been applied to the determination of lead in Bangladeshi fish samples. The slurry sampling conditions, such as slurry stabilizing agent, slurry concentration, pyrolysis temperature for the slurrie
Determination of zinc in biological materials by sequential metal vapor elution analysis (column temperature; >115OK) with hydrogen carrier gas and an atomic absorption detector (AA) is described. The column was a molybdenum tube inserted a tungsten coil. Zinc was completely separated from Al, Ca, C
An automated robotic method for sample pretreatment was developed for copper determination in environmental and biological materials at the low microgram per gram level. The robotic station performing the preliminary operations as weighing of sample and acid digestion was coupled with the flowinject
ment factor of 47 in comparison with direct injection of 45 lL aqueous solution was obtained at a sampling frequency of 31 h β«1Χβ¬ . The detection limit (3 r) was found to be 2.8 ng L β«1Χβ¬ . The precision was 3% relative standard deviation (RSD) at the 0.2 lg L β«1Χβ¬ level (n β«Χ‘β¬ 11). The results obta