✦ Synopsis


Abstract

Purpose:

To evaluate the effects of drift in scanner hardware and inter‐scanner variability on global and regional brain volume changes using longitudinal data obtained on two scanners of the exact same model.

Materials and Methods:

A total of 208 subjects (70 females and 138 males, mean age = 56.5 ± 10.1 years) were scanned twice, at an interval of approximately 2 years (mean interval = 1.9 ± 0.15 years), using two 3.0 Tesla (T) scanners of the exact same model. The subjects were divided into four groups according to the combination of scanners used. The effects of scanner drift and inter‐scanner variability on global and regional brain volume changes were evaluated using voxel‐based morphometry (VBM) and structural image evaluation using normalization of atrophy (SIENA).

Results:

The mean whole brain volume calculated using VBM increased by 17.2 ± 29.3 mL, and the changes were significantly different between groups. The mean percentage brain volume change (PBVC) calculated using SIENA was −0.46 ± 0.71 %, and did not differ between groups. VBM showed several regions with a significant increase or decrease in volume; however, SIENA showed that edge displacement values were negative in most regions.

Conclusion:

Even with scanners of the exact same model, scanner drift, and inter‐scanner variability can cancel out actual longitudinal brain volume change. SIENA, which corrects for differences in imaging geometry using the outer skull surface for both time points, can reduce the effects of scanner drift and inter‐scanner variability on longitudinal morphometric results. J. Magn. Reson. Imaging 2011;. © 2011 Wiley‐Liss, Inc.


📜 SIMILAR VOLUMES


Effect of scanner in longitudinal diffus
✍ Hidemasa Takao; Naoto Hayashi; Hiroyuki Kabasawa; Kuni Ohtomo 📂 Article 📅 2011 🏛 John Wiley and Sons 🌐 English ⚖ 750 KB

## Abstract The purpose of this study was to evaluate the effects of longitudinal drift in scanner hardware, inter‐scanner variability (bias), and scanner upgrade on longitudinal changes in global and regional diffusion properties using longitudinal data obtained on two scanners of the exact same m

Longitudinal change in forced expiratory
✍ Jay D. Pearson; Stephanie Y. Kao; Larry J. Brant; E. Jeffrey Metter; Melvyn S. T 📂 Article 📅 1998 🏛 John Wiley and Sons 🌐 English ⚖ 204 KB 👁 2 views

Studies of age-associated changes in forced expiratory volume at one second (FEV 1 ) have varied in the degree of screening for health and smoking, and have not examined age differences in variability in FEV 1 . Longitudinal rates of change and variability in FEV 1 among healthy lifetimenon-smoking

Longitudinal analysis of cognitive perfo
✍ C. Delaloye; G. Moy; F. de Bilbao; K. Weber; S. Baudois; S. Haller; A. Xekardaki 📂 Article 📅 2011 🏛 John Wiley and Sons 🌐 English ⚖ 365 KB

## Abstract ## Objectives Cross‐sectional studies in bipolar disorder (BD) suggested the presence of cognitive deficits and subtle magnetic resonance imaging (MRI) changes in limbic areas that may persist at euthymic stages. Whether or not cognitive and MRI changes represent stable attributes of B

Effects of brain-derived neurotrophic fa
✍ P. Cédric M.P. Koolschijn; Neeltje E.M. van Haren; Steven C. Bakker; Mechteld L. 📂 Article 📅 2010 🏛 John Wiley and Sons 🌐 English ⚖ 153 KB

## Abstract A functional polymorphism of the brain‐derived neurotrophic factor (BDNF) gene (Val66Met) has been associated with the risk for schizophrenia and volume differences in the hippocampus. However, little is known about the association between progressive brain volume change in schizophreni