Design and experimental demonstration of
โ
X.H. Li; X.J. Du; M. Qiu; Y.W. Ma; L.Y. Xiao
๐
Article
๐
2007
๐
Elsevier Science
๐
English
โ 141 KB
Test magnet for magnetic resonance imaging (MRI) applications using MgB 2 /Fe composite wire is designed and experimentally demonstrated. The magnetizing and screening effects of the Fe sheath, the cooling conditions and the field uniformity are considered carefully during the magnet designing. The