Inductively coupled plasma atomic emission spectrometry in combination with extraction chromatography was applied to the determination of trace amounts of zirconium and hafnium in high-purity scandium ozide. The chromatographic column separation procedure was carried out with 1-phenyl-3-methyl-4-ben
Determination of trace inpurities high-purity aluminum oxide by inductively-coupled plasma atomic emission-spectrometry
โ Scribed by T. Ishizuka; Y. Uwamino; A. Tsuge; T. Kamiyanagi
- Publisher
- Elsevier Science
- Year
- 1984
- Tongue
- English
- Weight
- 426 KB
- Volume
- 161
- Category
- Article
- ISSN
- 0003-2670
No coin nor oath required. For personal study only.
โฆ Synopsis
Trace impurities (Ca, Cu, Fe, Mg, Mn, Na and Si) in 99.99% aluminum oxide were determined by inductively-coupled plasma atomic emission spectrometry (i.c.p.la.e.6.). The sample was fused with lithium carbonate/boric acid to determine Ca, Cu, Mg, Mn, Na and Si or dissolved in phosphoric/sulphuric acids to determine iron. Matrix effects on the calibration graphs for each element were studied: for accurate determinations, calibration solutions must contain lithium and boric acid (or aluminum and phosphoric and sulfuric acids for iron).
๐ SIMILAR VOLUMES
An analytical method for the simultaneous determination of some trace elements (Au, Fe, Mg, Li, Sr, Zn) in human serum by inductively coupled plasma atomic emission spectrometry (ICP-AES) with flow injection is described. Physical interference caused by the change of sample viscosity is discussed. W
Fuel science and technology (fundamental science, analysis, instrumentation) 95lO5750 Datarmlnation of chlorine in coal by indirect induc. tiveiy coupled plasma-atomic emission spectromatry Liu, D. and Zhang, Y. Fenri Hua\*ue, 1996, 24, (2), p. 244. (In Chinese) Reports on the analysis of ten coal s
A desolvation device was connected to a microcolumn (40 mm X 3.0 mm i.d.) in which 8-hydroxyquinoline-5sulphonic acid was immobilized on the substrate of active carbon-silica gel. The analyte and 1.0 M sodium acetate-acetic acid buffer solution (pH 5.6) were mined and pumped through the microcolumn.