## Abstract Salicylate is a hydrolysis product of salicylateβcontaining drugs (such as aspirin) in patients' blood. Monitoring of this ion in blood is helpful for diagnosing of overdosage of these drugs. The present paper describes an ion chromatography (IC) method developed for determination of to
Determination of Iodide by Ion Chromatography with Series Bulk Acoustic Wave Detection
β Scribed by Xiaorong Yang; Po Chen; Wanzhi Wei; Shouzhuo Yao
- Publisher
- Elsevier Science
- Year
- 1998
- Tongue
- English
- Weight
- 91 KB
- Volume
- 58
- Category
- Article
- ISSN
- 0026-265X
No coin nor oath required. For personal study only.
β¦ Synopsis
A feasible chromatographic method for the determination of iodide using ion chromatography (IC) with series bulk acoustic wave (SBAW) detection has been developed. The advantages of the good selectivity of ion chromatography and the highly sensitive response of SBAW have been combined to improve detection limit, linearity, and reproducibility. The detection limit has been estimated to be 5 ng/ml. For the IC analysis a 12.5-mm polymethacrylate resin column incorporating a quaternary base was used, with the mobile phase being 5.0 mM phthalic acid. The method has been applied to the determination of iodide in some dairy products and table salt.
π SIMILAR VOLUMES
Ion chromatography with electrochemical detection was applied to the determination of bromide in snow samples. Greater reproducibility and increased sensitivity was obtained by coating the silver electrode with a highly insoluble AgI surface. Without preconcentration, a detection limit of less than