𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Determination of Cadmium in Wines Using Graphite Furnace Atomic Absorption Spectrometry with Zeeman Background Correction

✍ Scribed by James Jaganathan; Alan L. Reisig; Sumer M. Dugar


Publisher
Elsevier Science
Year
1997
Tongue
English
Weight
77 KB
Volume
56
Category
Article
ISSN
0026-265X

No coin nor oath required. For personal study only.

✦ Synopsis


A graphite furnace atomic absorption spectrometry method with Zeeman background correction for the direct determination of cadmium in grape wines is described. Use of ammonium dihydrogen phosphate, magnesium nitrate, and palladium nitrate as matrix modifiers was investigated. Palladium nitrate with 10% nitric acid was the most satisfactory matrix modifier. The methodological detection limit, calculated as 3s/slope of the calibration curve, was 0.03 mg/liter for cadmium in wine. The sensitivity, calculated as the characteristic mass, was 0.4 pg. The use of palladium nitrate as a matrix modifier also results in excellent reproducibility (Γ΅10% RSD) at such low levels. This method is readily applicable to the determination of cadmium in wines without any sample preparation.


πŸ“œ SIMILAR VOLUMES


Selenium Determination in Biological Flu
✍ Natalia Campillo; Pilar ViΓ±as; Ignacio LΓ³pez-Garcı́a; Manuel HernΓ‘ndez-CΓ³rdoba πŸ“‚ Article πŸ“… 2000 πŸ› Elsevier Science 🌐 English βš– 67 KB

Procedures for the direct determination of total selenium in urine, serum, and blood using electrothermal atomic absorption spectrometry are presented. In the selected experimental conditions, Zeeman correction is mandatory to compensate for the high background signals. The sample diluted and contai

Atomization of tin in saline water media
✍ Etsuro Iwamoto; Hiromichi Shimazu; Kayoko Yokota; Takahiro Kumamaru πŸ“‚ Article πŸ“… 1993 πŸ› Elsevier Science 🌐 English βš– 415 KB

The direct atonuzatlon of tm m sahne water media m graphite furnace atormc absorption spectrometry with a tungsten-coated tube was investigated The combmed use of a tungsten-coated non-pyrolytic graphite tube and palladium as a chenucal modlfier ehnunated sahmty m doubly diluted sea water without a

Determination of boron in iron- and nick
✍ Yongming Liu; Benling Gong; Yuli Xu; Zhuanhe Li; Tiezheng Lin πŸ“‚ Article πŸ“… 1994 πŸ› Elsevier Science 🌐 English βš– 320 KB

A chemical modifier composed of nickel and zirconium salts and treatment of the graphite tube with zirconium solution were studied for the determination of boron in cast iron, Fe-, Ni-and Fe-Ni-based alloys. A combination of 3000 mg 1-l nickel and 1000 mg I-' zirconium as chemical modifier and soaki

Determination of selenium and tellurium
✍ Benling Gong; Yongming Liu; Zhuanhe Li; Tiezheng Lin πŸ“‚ Article πŸ“… 1995 πŸ› Elsevier Science 🌐 English βš– 475 KB

A nickel/palladium matrix modifier was proposed for the determination of selenium and tellurium in nickel and nickel/iron-based alloys by graphite furnace atomic absorption spectrometry without preseparation of the matrix. The effect of this matrix modifier was studied and it was found that a compos